Parameters
16 different signals, including Time and Ratio (derived parameter) can be acquired simultaneously
in logarithmic and linear modes, giving up to 32 acquisition parameters from a possible 62.
5 different signals available from each detector:
• Integral linear
• Integral logarithmic
• Peak linear
• Peak logarithmic
• Time of Flight linear.
Regions
Up to 32 gates and 256 analysis regions are available for gating, analysis, and autogating per
protocol. Up to 8 of those regions can be used to create any one gate. The following types of regions
are available for gating and analysis:
• Linear
• Rectangular
• Quadrant
• Polygonal
• Elliptical
• Contour.
Listmode Analysis
The instrument can store up to 16 parameters including TIME, and RATIO as listmode data. The
amount of available random access memory (RAM) in the computer determines the maximum size
of a listmode file. See
Locked Protocols
User defined protocols can be locked or unlocked by a System Administrator and password
protected to prevent inadvertent modifications. The optional application software
(e.g. Navios tetra) uses locked protocols that cannot be unlocked by a System Administrator . The
following software screens are disabled when using locked protocols. You cannot make changes to
the locked features on these screens.
Dotplot Properties
Region Properties
Create / Modify Gates
Color Precedence
Color Blend
Cytometer Control
773232AH
RAM (Random Access
(QC product selection and export region statistics are not locked)
Memory).
Specifications
4
Software Specifications
4-7