Cytometer; Flow Cell; Flow Rate; Sheath Consumption - Beckman Coulter Navios Instructions For Use Manual

Flow cytometer
Table of Contents

Advertisement

Specifications
Instrument Specifications And Characteristics

Cytometer

Flow Cell

Sensing area: BioSense 150 μm x 460 rectangular channel with an integral lens, mounted with a
vertical (upward) flow path. See

Flow Rate

Continuous pressure is applied to the sample tube. The amount of pressure depends on the flow rate
you specify:
• Low approximately 10 uL/min
• Medium approximately 30 uL/min
• High approximately 60 uL/min

Sheath Consumption

• 780mL / hour (acquisition)
• 0mL / hour (standby)

Lasers

• Solid-state, software controlled, 22 mW, blue laser operating at 488 nm and
• Solid-state, software controlled, 25 mW, red laser operating at 638 nm.

Optional third laser:

• Solid-state, software controlled, 40 mW, violet laser operating at 405 nm.

Laser Power Monitoring

Laser power is monitored by each laser individually. If the laser power deviated more than ±10% a
Laser Power Error is displayed on the
measures rectify the fault. Follow the instructions in the Troubleshooting section for handling this
error.

Beam-Shaping Optics

Cross cylindrical lenses 10 mm by 80 mm.

Blue Laser Beam Spot Size

An elliptical spot 10 μm high by 84 μm wide.
4-2
Figure
3.1.
Status
Bar. The system will not run a sample until corrective
773232AH