Specifications
Instrument Specifications And Characteristics
Cytometer
Flow Cell
Sensing area: BioSense 150 μm x 460 rectangular channel with an integral lens, mounted with a
vertical (upward) flow path. See
Flow Rate
Continuous pressure is applied to the sample tube. The amount of pressure depends on the flow rate
you specify:
• Low approximately 10 uL/min
• Medium approximately 30 uL/min
• High approximately 60 uL/min
Sheath Consumption
• 780mL / hour (acquisition)
• 0mL / hour (standby)
Lasers
• Solid-state, software controlled, 22 mW, blue laser operating at 488 nm and
• Solid-state, software controlled, 25 mW, red laser operating at 638 nm.
Optional third laser:
• Solid-state, software controlled, 40 mW, violet laser operating at 405 nm.
Laser Power Monitoring
Laser power is monitored by each laser individually. If the laser power deviated more than ±10% a
Laser Power Error is displayed on the
measures rectify the fault. Follow the instructions in the Troubleshooting section for handling this
error.
Beam-Shaping Optics
Cross cylindrical lenses 10 mm by 80 mm.
Blue Laser Beam Spot Size
An elliptical spot 10 μm high by 84 μm wide.
4-2
Figure
3.1.
Status
Bar. The system will not run a sample until corrective
773232AH