3.0 3216 Controller
to select OFF, which is below the Zero value.
This Process is repeated for each of the 8 segments of the program.
If not all the segments are used for a program, the Ramp & Dwell of each of the
subsequent segments should be set to OFF.
A program will end in one of two ways, either revert to the control setpoint or dwell at
the temperature set in the last segment used. When a program finishes on a dwell and
the dwell time expires the temperature will revert to the control setpoint.
Note: Before operating a program ensure that the control setpoint is set to Zero to
avoid unexpected heating at the end of the program.
3.7.9
Running a Program
If using the 3216P5 controller press scroll
Select the required Program Number before operating a Program.
Table below shows the key presses to operate a program.
Operation
To RUN a program
To HOLD a program
To RESET a program Press and hold
To RESET a
program after it has
completed
3.7.10
Program Status
While the program is operating in level 1 or level 2, the home display shows two values
at any one time:
3.7.11
Process Value
The upper display shows the current temperature of the product.
18
Action
Press and quickly release
+
Press and quickly release
+
more than 1 second
Program Ended
Press and hold
more than 1 second or press
and quickly release Ack
to display PROG <PROGRAM NUMBER>.
Indication
Indicator – RUN = ON
Scrolling Display – Current
Program state
Indicator – RUN = Flashing
Scrolling Display – Program Hold
Indicator – RUN = OFF
+
for
Scrolling Display - None
Indicator – RUN = OFF
Scrolling Text – Program End
Indicator – RUN = OFF
+
for
Scrolling Display - None