Newport NPC1USB User Manual

Voltage piezo usb amplifier
Hide thumbs Also See for NPC1USB:
Table of Contents

Advertisement

Quick Links

NPC1USB
Voltage Piezo
USB Amplifier
User's Manual
V1.0.x

Advertisement

Table of Contents
loading

Summary of Contents for Newport NPC1USB

  • Page 1 NPC1USB Voltage Piezo USB Amplifier User’s Manual V1.0.x...
  • Page 2 Voltage Piezo USB Amplifier Warranty Newport Corporation warrants that this product will be free from defects in material and workmanship and will comply with Newport’s published specifications at the time of sale for a period of one year from date of shipment. If found to be defective during the warranty period, the product will either be repaired or replaced at Newport's option.
  • Page 3: Table Of Contents

    EU Declaration of Conformity ..................1.0 Introduction ....................1 Scope of the Manual ......................1 Certification of Newport Corporation ................1 Purchased Part Package ...................... 2 Instructions for Using Piezo Electrical Elements and Power Supplies ......2 Safety Instructions ......................3 1.5.1...
  • Page 4 NPC1USB Voltage Piezo USB Amplifier 5.1.2 Decimal Separator ....................13 5.1.3 Command Terminator ..................13 5.1.4 Command Execution Time ................13 5.1.5 Command Set ..................... 14 6.0 Command Overview ................... 15 ID — Set/Get Stage Identifier ..................16 MM – Enter/Leave DISABLE State ................17 OR —...
  • Page 5: Eu Declaration Of Conformity

    NPC1USB Voltage Piezo USB Amplifier EU Declaration of Conformity EDH0403En1010 — 07/17...
  • Page 6 NPC1USB Voltage Piezo USB Amplifier EDH0403En1010 — 07/17...
  • Page 7: Introduction

    Scope of the Manual This manual describes the piezo amplifier NPC1USB from Newport Corporation. If you have any problems please contact the manufacturer of the system:Newport Corporation, 1791 Irvine Ave. Irvine, CA 92606. Phone: (877) 835-9620 Certification of Newport Corporation Newport Corporation works in according to an ISO 9001:2008 certified quality management system.
  • Page 8: Purchased Part Package

    • Please use only original parts from Newport Corporation. • Please contact Newport Corporation or your local representative, if there are any problems with your actuator or power supply. EDH0403En1010 — 07/17...
  • Page 9: Safety Instructions

    NPC1USB Voltage Piezo USB Amplifier CAUTION Shock forces may damage the built-in ceramic elements. Please avoid such forces, and handle the units with care, otherwise the guarantee is voided. Safety Instructions 1.5.1 General Warning or Caution Figure 1: General Warning or Caution Symbol.
  • Page 10: Warnings And Cautions

    For information about where the user can drop off the waste equipment for recycling, please contact your local Newport representative. Control of Hazardous Substances RoHS Compliant Symbol This label indicates the products comply with the EU Directive 2011/65/EC that restricts the content of six hazardous chemicals.
  • Page 11: Warnings And Cautions

    (e.g. actuators) can work under other conditions. • Please only use original parts from Newport Corporation. • Newport Corporation does not provide any warranty for damages or malfunctions caused by additional parts not supplied by Newport Corporation. Additional cables or connectors will change the calibration and other specified data.
  • Page 12: Installation, Power Supply

    CAUTION If the voltage amplifier emits smoke, excess heat, or unusual smells, immediately turn off the power switch and unplug the power plug from the outlet. Then contact Newport Corporation for technical services. EDH0403En1010 — 07/17...
  • Page 13: Maintenance And Inspection

    NPC1USB Voltage Piezo USB Amplifier Maintenance and Inspection CAUTION! • When cleaning the exterior box of the voltage amplifier, first turn off the power switch and unplug the power plug. Failure to follow these procedures may result in a fire or electrical shock.
  • Page 14: Npc1Usb Definition

    The amplifier can be controlled by an applied analogue voltage signal from 0 V to +10 V. Monitor Output The output voltage signal corresponds to the position of the piezo actuating system with the applied voltage signal (OPEN LOOP mode only for NPC1USB). PC Interface USB 2.0 Hyper Terminal Program This software enables the control of the piezo actuating system by using the USB interface.
  • Page 15: User Elements/Connections

    3.3.1 Block Diagram The chart below shows the block diagram of the NPC1USB. With exception of the display and the processor, the circuit is a similar of the NPC amplifier series. NOTE The travel capability of actuators or stages is limited to 80% due to the limited voltage range of the NPC1USB.
  • Page 16: Amplifier Without Integrated Feedback Controller

    Newport Corporation and will require additional effort and costs. Please contact our sales team for more information. Initiation Please follow all safety instructions given in this user manual and the NPC1USB Quick Start Guide. 3.4.1 Main Supply Voltage The NPC1USB amplifier requires a main supply voltage of 24 VDC.
  • Page 17: Handling

    Handling Controlling via Interface The amplifier NPC1USB is controlled via an USB 2.0 interface. Therefore, signal parameters and information about the actuator position or status can be adjusted or settled directly in the computer interface. Furthermore, the system settings of the amplifier unit can also be changed by using the interface.
  • Page 18: State Diagram

    Voltage Piezo USB Amplifier State Diagram For a safe and consistent operation, the NPC1USB Uses 6 different operation states: Not referenced, Configuration, Homing, Ready, Disable and Moving. In each state, only specific commands are accepted by the NPC1USB. It is important to understand the state diagram below and which commands and actions cause transition between different states.
  • Page 19: Command Syntax

    READY state. The process for homing, and which signals are looked for during homing, can be defined with the HT command. Command Syntax The NPC1USB is a command driven controller. The general format of a command is a two letter ASCII character preceded and followed by parameters specific to the command: Command format: Optional or required controller address.
  • Page 20: Command Set

    For example, a 1VA10 sets the velocity of the controller #1 to 10 units/second. A “1VA?” sends the response “1VA10”. Not every command can be executed in all states of the NPC1USB and some commands have different meaning in different states. It is therefore important to understand the state diagram of the controller, see chapter 5.0: “State Diagram”.
  • Page 21: Command Overview

    NPC1USB Voltage Piezo USB Amplifier Command Overview Not Ref. Config. Disable Ready Motion Description – Set/Get stage identifier    – – – Enter/Leave DISABLE state   – – – – Enable output stage and set initial voltage ...
  • Page 22: Id - Set/Get Stage Identifier

    Execution not allowed in NOT REFERENCED state. — Execution not allowed in HOMING state. — Execution not allowed in MOVING state. Rel. Commands None. Example 1ID? Get stage identifier for controller #1. 1IDNPC1USB Controller returns stage identifier: NPC1USB. EDH0403En1010 — 07/17...
  • Page 23: Mm - Enter/Leave Disable State

    NPC1USB Voltage Piezo USB Amplifier MM – Enter/Leave DISABLE State Usage Not Ref. Config. Disable Ready Motion – – –   Syntax xxMMnn or xxMM? Parameters Description xx [int] — Controller address. nn [float] — Range — 1 to 31.
  • Page 24: Or - Enable Output Stage And Set Initial Voltage

    NPC1USB Voltage Piezo USB Amplifier OR — Enable Output Stage and Set Initial Voltage Usage Not Ref. Config. Disable Ready Motion – – – –  Syntax xxOR Parameters Description xx [int] — Controller address. Range — 1 to 31.
  • Page 25: Pa - Move Absolute

    NPC1USB Voltage Piezo USB Amplifier PA — Move Absolute Usage Not Ref. Config. Disable Ready Motion – – – –  Syntax xxPAnn or xxPA? Parameters Description xx [int] — Controller address. nn [float] — Actuator voltage. Range — 1 to 31.
  • Page 26: Pr - Move Relative

    NPC1USB Voltage Piezo USB Amplifier PR — Move Relative Usage Not Ref. Config. Disable Ready Motion – – – –  Syntax xxPRnn or xxPR? Parameters Description xx [int] — Controller address. nn [float] — Range — 1 to 31.
  • Page 27: Pw - Enter/Leave Configuration State

    NPC1USB Voltage Piezo USB Amplifier PW — Enter/Leave CONFIGURATION State Usage Not Ref. Config. Disable Ready Motion – – –   Syntax xxPWnn or xxPW? Parameters Description xx [int] — Controller address. nn [float] — Range — 1 to 31.
  • Page 28: Rs - Reset Controller

    NPC1USB Voltage Piezo USB Amplifier RS — Reset Controller Usage Not Ref. Config. Disable Ready Motion – –    Syntax xxRS or RS Parameters Description xx [int] — Controller address. Range — 1 to 31. Units — None.
  • Page 29: Rs## - Reset Controller's Address To 1

    The RS## command resets the controller’s address to 1. This address needs to be different for each NPC1USB when connected on a RS-485 communication network. Should the address change be saved in the device please store it with the sequence 1PW1, 1PW0.
  • Page 30: Sa - Set/Get Controller's Rs-485 Address

    Out of range:Error C. Description The SA command sets the controller’s RS-485 address. Newport Corporation recommends using the supplied utility software for all controller configurations. The SA command is of practical use only when not using this software. Returns If the sign “?” takes place of nn, this command returns the current programmed value.
  • Page 31: Se - Do Nothing, No Error For Compatibility

    NPC1USB Voltage Piezo USB Amplifier SE — Do Nothing, No Error for Compatibility Usage Not Ref. Config. Disable Ready Motion – – – –  Syntax xxSEnn, or xxSE? or SE Parameters Description xx [int] — Controller address. nn [float] —...
  • Page 32: Sl - Set/Get Negative Software Limit

    NPC1USB Voltage Piezo USB Amplifier SL — Set/Get Negative Software Limit Usage Not Ref. Config. Disable Ready Motion – –    Syntax xxSLnn or xxSL? Parameters Description xx [int] — Controller address. nn [float] — Negative software limit.
  • Page 33: Sr - Set/Get Positive Software Limit

    NPC1USB Voltage Piezo USB Amplifier SR — Set/Get Positive Software Limit Usage Not Ref. Config. Disable Ready Motion – –    Syntax xxSRnn or xxSR? Parameters Description xx [int] — Controller address. nn [float] — Positive software limit.
  • Page 34: St - Stop Motion

    NPC1USB Voltage Piezo USB Amplifier ST — Stop Motion Usage Not Ref. Config. Disable Ready Motion – – – –  Syntax xxST Parameters Description xx [int] — Controller address. Range — 1 to 31. Units — None. Defaults Missing: Error B.
  • Page 35: Tb - Get Command Error String

    NPC1USB Voltage Piezo USB Amplifier TB — Get Command Error String Usage Not Ref. Config. Disable Ready Motion      Syntax xxTBnn Parameters Description xx [int] — Controller address. nn [char] — Error code (refer to TE command).
  • Page 36: Te - Get Last Command Error

    NPC1USB Voltage Piezo USB Amplifier TE — Get Last Command Error Usage Not Ref. Config. Disable Ready Motion      Syntax xxTE Parameters Description xx [int] — Controller address. Range — 1 to 31. Units — None.
  • Page 37: Th - Set/ Get Point Position

    NPC1USB Voltage Piezo USB Amplifier TH — Set/ Get Point Position Usage Not Ref. Config. Disable Ready Motion      Syntax xxTH Parameters Description xx [int] — Controller address. Range — 1 to 31. Units — None.
  • Page 38: Tp - Same As Th For Compatibility

    NPC1USB Voltage Piezo USB Amplifier TP — Same as TH for Compatibility Usage Not Ref. Config. Disable Ready Motion      Syntax xxTP Parameters Description xx [int] — Controller address. Range — 1 to 31. Units —...
  • Page 39: Ts - Get Positioner Error And Controller State

    NPC1USB Voltage Piezo USB Amplifier TS — Get Positioner Error and Controller State Usage Not Ref. Config. Disable Ready Motion      Syntax xxTS Parameters Description xx [int] — Controller address. Range — 1 to 31. Units —...
  • Page 40 NPC1USB Voltage Piezo USB Amplifier Controller States (ef) ef [HEX] ef [BIN] Description 1010 NOT REFERENCED from reset 1011 NOT REFERENCED from HOMING 1100 NOT REFERENCED from CONFIGURATION 1101 NOT REFERENCED from DISABLE 1110 NOT REFERENCED from READY 1111 NOT REFERENCED from MOVING...
  • Page 41: Va - Set/Get Slow Rate

    NPC1USB Voltage Piezo USB Amplifier VA — Set/Get Slow Rate Usage Not Ref. Config. Disable Ready Motion – –    Syntax xxVAnn or xxVA? Parameters Description xx [int] — Controller address. nn [float] — Slew rate. Range —...
  • Page 42: Ve - Get Controller Revision Information

    NPC1USB Voltage Piezo USB Amplifier VE — Get Controller Revision Information Usage Not Ref. Config. Disable Ready Motion      Syntax xxVE Parameters Description xx [int] — Controller address. Range — 1 to 31. Units — None.
  • Page 43: Zt - Get All Axis Parameters

    Description The ZT command returns the list of all current configuration parameters. The ZT command allows a quick review of all current stage parameter and simplifies the configuration of non-Newport stages(e.g. by using Hyper Terminal file transfer). Returns All parameters.
  • Page 44: Pinouts

    NPC1USB Voltage Piezo USB Amplifier Pinouts 24 VDC IN and OUT (Female ø 2.1 Ø 5.5 11 mm) Pin # Description Center +24 VDC Outer Ground USB Connector (USB Mini-B Receptacle) Pin # Description +5 VDC IN, Do not connect if comm connector is used Data –...
  • Page 45: Electrical Specifications

    NPC1USB Voltage Piezo USB Amplifier Electrical Specifications Parameter Specification Remarks Channels Output voltage 0…+130 Output current [mA] continuous Output connector JST S4B-XH-A Inside Adapter cable SUB-D15M Outside Interface USB 2.0 Resolution Main supply 24 V ±10%/0.2 A Main supply connector 2.1 mm Barrel type...
  • Page 46 NPC1USB Voltage Piezo USB Amplifier EDH0403En1010 — 07/17...
  • Page 47: Service Form

    NPC1USB Voltage Piezo USB Amplifier Service Form Your Local Representative Tel.: __________________ Fax: ___________________ Name: _________________________________________________ Return authorization #: ____________________________________ (Please obtain prior to return of item) Company:_______________________________________________ Address: ________________________________________________ Date: __________________________________________________ Country: ________________________________________________ Phone Number: __________________________________________ P.O. Number: ____________________________________________...
  • Page 48 Visit Newport Online at: www.newport.com North America & Asia Europe Newport Corporation MICRO-CONTROLE Spectra-Physics S.A.S 1791 Deere Ave. 9, rue du Bois Sauvage Irvine, CA 92606, USA 91055 Évry CEDEX France Sales Tel.: (800) 222-6440 Sales e-mail: sales@newport.com Tel.: +33 (0)1.60.91.68.68 Technical Support e-mail: france@newport.com...

Table of Contents