Page 1
NPC1USB Voltage Piezo USB Amplifier User’s Manual V1.0.x...
Page 2
Voltage Piezo USB Amplifier Warranty Newport Corporation warrants that this product will be free from defects in material and workmanship and will comply with Newport’s published specifications at the time of sale for a period of one year from date of shipment. If found to be defective during the warranty period, the product will either be repaired or replaced at Newport's option.
EU Declaration of Conformity ..................1.0 Introduction ....................1 Scope of the Manual ......................1 Certification of Newport Corporation ................1 Purchased Part Package ...................... 2 Instructions for Using Piezo Electrical Elements and Power Supplies ......2 Safety Instructions ......................3 1.5.1...
Page 4
NPC1USB Voltage Piezo USB Amplifier 5.1.2 Decimal Separator ....................13 5.1.3 Command Terminator ..................13 5.1.4 Command Execution Time ................13 5.1.5 Command Set ..................... 14 6.0 Command Overview ................... 15 ID — Set/Get Stage Identifier ..................16 MM – Enter/Leave DISABLE State ................17 OR —...
Scope of the Manual This manual describes the piezo amplifier NPC1USB from Newport Corporation. If you have any problems please contact the manufacturer of the system:Newport Corporation, 1791 Irvine Ave. Irvine, CA 92606. Phone: (877) 835-9620 Certification of Newport Corporation Newport Corporation works in according to an ISO 9001:2008 certified quality management system.
• Please use only original parts from Newport Corporation. • Please contact Newport Corporation or your local representative, if there are any problems with your actuator or power supply. EDH0403En1010 — 07/17...
NPC1USB Voltage Piezo USB Amplifier CAUTION Shock forces may damage the built-in ceramic elements. Please avoid such forces, and handle the units with care, otherwise the guarantee is voided. Safety Instructions 1.5.1 General Warning or Caution Figure 1: General Warning or Caution Symbol.
For information about where the user can drop off the waste equipment for recycling, please contact your local Newport representative. Control of Hazardous Substances RoHS Compliant Symbol This label indicates the products comply with the EU Directive 2011/65/EC that restricts the content of six hazardous chemicals.
(e.g. actuators) can work under other conditions. • Please only use original parts from Newport Corporation. • Newport Corporation does not provide any warranty for damages or malfunctions caused by additional parts not supplied by Newport Corporation. Additional cables or connectors will change the calibration and other specified data.
CAUTION If the voltage amplifier emits smoke, excess heat, or unusual smells, immediately turn off the power switch and unplug the power plug from the outlet. Then contact Newport Corporation for technical services. EDH0403En1010 — 07/17...
NPC1USB Voltage Piezo USB Amplifier Maintenance and Inspection CAUTION! • When cleaning the exterior box of the voltage amplifier, first turn off the power switch and unplug the power plug. Failure to follow these procedures may result in a fire or electrical shock.
The amplifier can be controlled by an applied analogue voltage signal from 0 V to +10 V. Monitor Output The output voltage signal corresponds to the position of the piezo actuating system with the applied voltage signal (OPEN LOOP mode only for NPC1USB). PC Interface USB 2.0 Hyper Terminal Program This software enables the control of the piezo actuating system by using the USB interface.
3.3.1 Block Diagram The chart below shows the block diagram of the NPC1USB. With exception of the display and the processor, the circuit is a similar of the NPC amplifier series. NOTE The travel capability of actuators or stages is limited to 80% due to the limited voltage range of the NPC1USB.
Newport Corporation and will require additional effort and costs. Please contact our sales team for more information. Initiation Please follow all safety instructions given in this user manual and the NPC1USB Quick Start Guide. 3.4.1 Main Supply Voltage The NPC1USB amplifier requires a main supply voltage of 24 VDC.
Handling Controlling via Interface The amplifier NPC1USB is controlled via an USB 2.0 interface. Therefore, signal parameters and information about the actuator position or status can be adjusted or settled directly in the computer interface. Furthermore, the system settings of the amplifier unit can also be changed by using the interface.
Voltage Piezo USB Amplifier State Diagram For a safe and consistent operation, the NPC1USB Uses 6 different operation states: Not referenced, Configuration, Homing, Ready, Disable and Moving. In each state, only specific commands are accepted by the NPC1USB. It is important to understand the state diagram below and which commands and actions cause transition between different states.
READY state. The process for homing, and which signals are looked for during homing, can be defined with the HT command. Command Syntax The NPC1USB is a command driven controller. The general format of a command is a two letter ASCII character preceded and followed by parameters specific to the command: Command format: Optional or required controller address.
For example, a 1VA10 sets the velocity of the controller #1 to 10 units/second. A “1VA?” sends the response “1VA10”. Not every command can be executed in all states of the NPC1USB and some commands have different meaning in different states. It is therefore important to understand the state diagram of the controller, see chapter 5.0: “State Diagram”.
Execution not allowed in NOT REFERENCED state. — Execution not allowed in HOMING state. — Execution not allowed in MOVING state. Rel. Commands None. Example 1ID? Get stage identifier for controller #1. 1IDNPC1USB Controller returns stage identifier: NPC1USB. EDH0403En1010 — 07/17...
NPC1USB Voltage Piezo USB Amplifier MM – Enter/Leave DISABLE State Usage Not Ref. Config. Disable Ready Motion – – – Syntax xxMMnn or xxMM? Parameters Description xx [int] — Controller address. nn [float] — Range — 1 to 31.
NPC1USB Voltage Piezo USB Amplifier OR — Enable Output Stage and Set Initial Voltage Usage Not Ref. Config. Disable Ready Motion – – – – Syntax xxOR Parameters Description xx [int] — Controller address. Range — 1 to 31.
NPC1USB Voltage Piezo USB Amplifier PA — Move Absolute Usage Not Ref. Config. Disable Ready Motion – – – – Syntax xxPAnn or xxPA? Parameters Description xx [int] — Controller address. nn [float] — Actuator voltage. Range — 1 to 31.
NPC1USB Voltage Piezo USB Amplifier PR — Move Relative Usage Not Ref. Config. Disable Ready Motion – – – – Syntax xxPRnn or xxPR? Parameters Description xx [int] — Controller address. nn [float] — Range — 1 to 31.
NPC1USB Voltage Piezo USB Amplifier PW — Enter/Leave CONFIGURATION State Usage Not Ref. Config. Disable Ready Motion – – – Syntax xxPWnn or xxPW? Parameters Description xx [int] — Controller address. nn [float] — Range — 1 to 31.
NPC1USB Voltage Piezo USB Amplifier RS — Reset Controller Usage Not Ref. Config. Disable Ready Motion – – Syntax xxRS or RS Parameters Description xx [int] — Controller address. Range — 1 to 31. Units — None.
The RS## command resets the controller’s address to 1. This address needs to be different for each NPC1USB when connected on a RS-485 communication network. Should the address change be saved in the device please store it with the sequence 1PW1, 1PW0.
Out of range:Error C. Description The SA command sets the controller’s RS-485 address. Newport Corporation recommends using the supplied utility software for all controller configurations. The SA command is of practical use only when not using this software. Returns If the sign “?” takes place of nn, this command returns the current programmed value.
NPC1USB Voltage Piezo USB Amplifier SE — Do Nothing, No Error for Compatibility Usage Not Ref. Config. Disable Ready Motion – – – – Syntax xxSEnn, or xxSE? or SE Parameters Description xx [int] — Controller address. nn [float] —...
NPC1USB Voltage Piezo USB Amplifier ST — Stop Motion Usage Not Ref. Config. Disable Ready Motion – – – – Syntax xxST Parameters Description xx [int] — Controller address. Range — 1 to 31. Units — None. Defaults Missing: Error B.
NPC1USB Voltage Piezo USB Amplifier TE — Get Last Command Error Usage Not Ref. Config. Disable Ready Motion Syntax xxTE Parameters Description xx [int] — Controller address. Range — 1 to 31. Units — None.
NPC1USB Voltage Piezo USB Amplifier TH — Set/ Get Point Position Usage Not Ref. Config. Disable Ready Motion Syntax xxTH Parameters Description xx [int] — Controller address. Range — 1 to 31. Units — None.
NPC1USB Voltage Piezo USB Amplifier TP — Same as TH for Compatibility Usage Not Ref. Config. Disable Ready Motion Syntax xxTP Parameters Description xx [int] — Controller address. Range — 1 to 31. Units —...
NPC1USB Voltage Piezo USB Amplifier TS — Get Positioner Error and Controller State Usage Not Ref. Config. Disable Ready Motion Syntax xxTS Parameters Description xx [int] — Controller address. Range — 1 to 31. Units —...
Page 40
NPC1USB Voltage Piezo USB Amplifier Controller States (ef) ef [HEX] ef [BIN] Description 1010 NOT REFERENCED from reset 1011 NOT REFERENCED from HOMING 1100 NOT REFERENCED from CONFIGURATION 1101 NOT REFERENCED from DISABLE 1110 NOT REFERENCED from READY 1111 NOT REFERENCED from MOVING...
NPC1USB Voltage Piezo USB Amplifier VA — Set/Get Slow Rate Usage Not Ref. Config. Disable Ready Motion – – Syntax xxVAnn or xxVA? Parameters Description xx [int] — Controller address. nn [float] — Slew rate. Range —...
NPC1USB Voltage Piezo USB Amplifier VE — Get Controller Revision Information Usage Not Ref. Config. Disable Ready Motion Syntax xxVE Parameters Description xx [int] — Controller address. Range — 1 to 31. Units — None.
Description The ZT command returns the list of all current configuration parameters. The ZT command allows a quick review of all current stage parameter and simplifies the configuration of non-Newport stages(e.g. by using Hyper Terminal file transfer). Returns All parameters.
NPC1USB Voltage Piezo USB Amplifier Pinouts 24 VDC IN and OUT (Female ø 2.1 Ø 5.5 11 mm) Pin # Description Center +24 VDC Outer Ground USB Connector (USB Mini-B Receptacle) Pin # Description +5 VDC IN, Do not connect if comm connector is used Data –...
NPC1USB Voltage Piezo USB Amplifier Electrical Specifications Parameter Specification Remarks Channels Output voltage 0…+130 Output current [mA] continuous Output connector JST S4B-XH-A Inside Adapter cable SUB-D15M Outside Interface USB 2.0 Resolution Main supply 24 V ±10%/0.2 A Main supply connector 2.1 mm Barrel type...
Page 46
NPC1USB Voltage Piezo USB Amplifier EDH0403En1010 — 07/17...
NPC1USB Voltage Piezo USB Amplifier Service Form Your Local Representative Tel.: __________________ Fax: ___________________ Name: _________________________________________________ Return authorization #: ____________________________________ (Please obtain prior to return of item) Company:_______________________________________________ Address: ________________________________________________ Date: __________________________________________________ Country: ________________________________________________ Phone Number: __________________________________________ P.O. Number: ____________________________________________...
Page 48
Visit Newport Online at: www.newport.com North America & Asia Europe Newport Corporation MICRO-CONTROLE Spectra-Physics S.A.S 1791 Deere Ave. 9, rue du Bois Sauvage Irvine, CA 92606, USA 91055 Évry CEDEX France Sales Tel.: (800) 222-6440 Sales e-mail: sales@newport.com Tel.: +33 (0)1.60.91.68.68 Technical Support e-mail: france@newport.com...