Nozzle location (continued)
Provide easy access for viewing the display and programming via the hand
programmer.
Provide an environment suitable to the housing rating and materials of construction.
Provide a sunshield if the instrument will be mounted in direct sunlight.

Orientation in a vessel with obstructions

Polarization reference point
7ML19985KE03
polarization
axis
polarization
reference
point
display
SITRANS LR260 (HART) – INSTRUCTION MANUAL
ambient temperature
–40 °C to +80 °C (–40 °F to +176 °F)
process temperature (with FKM O-ring)
non-pressure version: –40 to +200 °C (–40 to +392 °F)
pressure rated version: –40 to +80 °C (–40 to +176 °F)
For best results on a vessel with
obstructions, orient the front or back of
the device toward the obstructions.
Page 15