1742/1746/1748
Users Manual
iFlex Probe Specifications
Flexible Current Probe Specifications
Measuring range
Weight
Probe Cable Length
Probe Cable Diameter
Minimum Bending Radius
Nondestructive current
Intrinsic Error at reference condition
Accuracy Logger + iFlex
Temperature Coefficient over
operating temperature range
Working Voltage
Output Cable length
Probe Cable Material
Coupling Material
Output Cable Material
Temperature, operating
Temperature, non-operating
Relative humidity, operating
60
i17XX-FLEX1.5KIP
1 A ac to 150 A ac
10 A ac to 1500 A ac
170 g (0.38 lb)
610 mm (24 in)
7.5 mm (0.3 in)
38 mm (1.5 in)
100 kA (50/60 Hz)
±0.7 % of reading [Reference Condition: Environmental: 23 °C ±5 °C, no external electrical/magnetic field, RH 65 %.
Primary conductor in center position]
±(1 % of reading + 0.02 % of range)
0.05 % of reading / °C (0.028 % of reading / °F)
1000 V CAT III, 600 V CAT IV
2.5 m (8.2 ft)
TPR
POM + ABS/PC
TPR/PVC
-25 °C to +70 °C (-13 °F to +158 °F) temperature of conductor under test shall not exceed 80 °C (176 °F)
-40 °C to +80 °C (-40 °F to +176°F)
IEC 60721-3-3: 3K6:
-25 °C to +30 °C (-13 °F to +86 °F): ≤100 %
40 °C (104 °F): 55 %
50 °C (122 °F): 35 %
i17XX-FLEX3KIP
3 A ac to 300 A ac
30 A ac to 3000 A ac
170 g (0.38 lb)
610 mm (24 in)
i17XX-FLEX6KIP
6 A ac to 600 A ac
60 A ac to 6000 A ac
190 g (0.42 lb)
915 mm (36 in)
±(1.5 % of reading + 0.03 % of range)
0.1 % of reading / °C (0.056 % of
reading / °F)