Wireless Blood Pressure Wrist Monitor (BP7) OWNER’S MANUAL Table of Contents INTRODUCTION ......................... 2 PACKAGE CONTENTS ......................2 INTENDED USE ........................... 2 BLOOD PRESSURE CLASSIFICATION FOR ADULTS ............2 CONTRAINDICATION ....................... 2 PARTS AND DISPLAY INDICATORS ..................3 SET UP REQUIREMENTS ......................3 SET UP PROCEDURES.......................
INTRODUCTION Thank you for selecting the iHealth Wireless Blood Pressure Wrist Monitor. The iHealth Wireless Blood Pressure Wrist Monitor is a fully automatic wrist cuff blood pressure monitor that uses the oscillometric principle to measure your blood pressure and pulse rate. The monitor works with your iOS devices to test, track and share vital blood pressure data.
PARTS AND DISPLAY INDICATORS SET UP REQUIREMENTS The iHealth Wireless Blood Pressure Wrist Monitor is designed to be used with the following iPod touch, iPhone and iPad models: iPod touch (4th generation) iPod touch (3rd generation) iPhone 4S iPhone 4...
Place your hand palm-side up in front of you and wrap the cuff around a bare wrist 1-2 cm above the wrist joint b. If the monitor is correctly placed, the iHealth logo will be facing upright. c. The cuff must not be too tight nor too loose.
Page 6
START/STOP button. The status indicator light will be green and the monitor will begin measurement. If measured successfully, the measurement will be stored in the iHealth Wireless Blood Pressure Wrist Monitor and the status indicator light will flash green. The reading is not visible until the next iOS connection.
SPECIFICATIONS 1. Product name: iHealth Wireless Blood Pressure Wrist Monitor 2. Model: BP7 3. Classification: Internally powered, Type BF applied part, IPX0, No AP or APG, Continuous operation 4. Machine size: approx. 2.8” x 2.9” x 0.7” (72mm × 74mm × 17.6mm) 5.
Page 8
d) Simultaneous use with other medical monitoring equipment on the same limb e) The blood circulation of the user needs to be checked This Wireless Blood Pressure Wrist Monitor is designed for adults and should never be used on infants, young children, pregnant or pre-eclamptic patients. Consult your physician before use on children.
: Note The battery has limited charge cycles and may eventually need to replaced by an iHealth service provider. Battery life and charge cycles vary by use and settings. Monitor Status Status Indicator Charging...
Retest when calm; avoid speaking or excited or nervous during movement during the test test Bluetooth connection Bluetooth connection Reset iOS device. Reset BP7 by unstable unsuccessful, BP7 is pressing the START/STOP button abnormal, or strong about 10s. Make sure the monitor...
Battery replacement should only be performed by a qualified iHealth technician. To do otherwise will void your warranty and possibly damage your unit. 11. Cuff replacement should only be performed by a qualified iHealth technician. To do otherwise will possibly damage your unit.
Apple Inc., registered in the U.S. and other countries. The Bluetooth® word mark and logos are registered trademarks owned by Bluetooth SIG, Inc. and any use of such marks by iHealth Lab Inc. is under license. Other trademarks and trade names are those of their respective owners.
harmful interference to radio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures: —Reorient or relocate the receiving antenna. —Increase the separation between the equipment and receiver.
Page 14
nearby electronic equipment. RF emissions The monitor is suitable for use in all establishments, Class B CISPR 11 including domestic establishments and those directly connected to the public low-voltage power supply network Harmonic emissions Class A that supplies buildings used for domestic purposes. IEC 61000-3-2 Voltage fluctuations/ flicker emissions...
Page 15
Guidance and manufacturer’s declaration - electromagnetic immunity The Wireless Blood Pressure Wrist Monitor is intended for use in the electromagnetic environment specified below. The customer or the user of the Wireless Blood Pressure Wrist Monitor should assure that it is used in such an environment. IMMUNITY IEC 60601test Compliance...
Page 16
above, the monitor should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating the monitor. b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m. Table 4 For ME EQUIPMENT and ME SYSTEMS that are not LIFE-SUPPORTING Recommended separation distances between...
Need help?
Do you have a question about the BP7 and is the answer not in the manual?