Control Panel - Biomega BioClave Operation Manual

Autoclave
Table of Contents

Advertisement

APPENDIX 1
Water Properties/Characteristics:
DESCRIPTION
Evaporate residue
Silicium oxide sio
2
Iron
Cadmium
Lead
Rest of heavy metals, excluding
iron, cadmium, lead
Chloride
Phosphates
Conductivity (at 20 ) 。
pH value
Appearance
Hardness
Instructions manual
WATER
CONDENSATE
。10 mg/l
。1.0 mg/kg
。0.1 mg/kg
。1 mg/l
。0.1 mg/kg
。0.2 mg/l
。0.05 mg/kg
。0.005 mg/l
。0.1 mg/kg
。0.05 mg/l
。0.1 mg/l
。0.1 mg/kg
。2 mg/l
。0.1 mg/l
。0.1 mg/l
。0.5 mg/l
。 ヲ
15 s/cm
3 s/cm
5-7
5-7.5
Colorless, clean,
Colorless, clean,
without sediments
without sediments
。0.02 mmol/l
。0.02 mmol/l

5. Control panel

5.1 Pressure display window.
Displays the current pressure (PSI)
inside the chamber
5.2 Temperature display window.
Display the current temperature (°C)
inside the chamber
5.3 Timer display window
Displays the cycle state (see below)
or the remaining cycle time
5.4 TEMP button
Select the desired sterilization temperature
according to your instrument.
5.5 Sterilization Cycles
-
UNWRAPPED
- WRAPPED
-
LIQUID
- DRY ONLY
(Please see Appendix 2)
5.6 WATER indication lamp
This lamp will illuminate when the distilled water tank .
is low on water. The tank should then be filled (with distilled water).
5.7 FULL indication lamp
This lamp will illuminate when the used water tank is full. The used water
tank must be drained to continue operation.
5.8 START & STOP button
Once the parameters have been chosen, press this button to start the sterilization
cycle. To cancel or stop the cycle, press and hold this button for 3 sec.
Time Display
Ld
Do
Po
HE
Time in Min.
PL
Ed
E1 - E9
5
1 8
Cycle
State
State
State description
The machine is prepared for sample loading
preheating / loading
Door is open
Door state
The machine is vacuuming during a cycle
vacuuming
The machine is in heating mode
heating
Time remaining in program
sterilization/drying
The machine is cooling and releasing steam.
exhaust
The cycle is completed, you can open the door
end
and take out the sterilized sample.
Error state
Error - Please see section 12
Instructions manual

Hide quick links:

Advertisement

Table of Contents
loading

Related Products for Biomega BioClave

Table of Contents