APPENDIX 4 Thank you for choosing the Biomega BioClave steam sterilizer. HYDRAULIC DRAWING Your steam sterilizer has been CE certified and designed with durability, reliability, and safety in Safety valve mind. It is your responsibility to install this instrument in conformance with local electrical...
1. General The sterilizer described in this manual is intended for the sterilization Pressure psi of research tools. It operates automatically with 134°C and 121°C LIQUID sterilization temperatures. This sterilizer is in compliance with the 134°C European Directive 93/42/CEE and it has been produced in accordance with the EN 13060.
APPENDIX 2 4. Installation DIAGRAMS OF THE STERILIZATION PROGRAMMES Ensure that the sterilizer is installed with 2.5in. (10cm) ventilation space on all sides of the sterilizer, and 5 in. (20cm) on top side. The clearance required to open the door is 15.5in. (40cm). PROGRAM TYPE The sterilizer should be placed on a level worktable.
5. Control panel APPENDIX 1 Water Properties/Characteristics: 5.1 Pressure display window. Displays the current pressure (PSI) inside the chamber 5.2 Temperature display window. Display the current temperature (°C) WATER CONDENSATE DESCRIPTION inside the chamber 。10 mg/l 。1.0 mg/kg 5.3 Timer display window Evaporate residue Displays the cycle state (see below) or the remaining cycle time...
For additional information on any of the error codes listed in section 6.1.1 Open the chamber door and remove all inner packing and accessories. 10, please contact your sales representative, or contact Biomega 6.1.2 Plug the power cord into the proper electrical outlet. (Please Research Products Service Department at 1-908-769-5555. Please check that the power source is in accordance to the electrical specifications of the machine (listed on the power label).
12. Error Codes * Verify all samples are not sealed, capped or closed. * Don't overload the trays above the stated limit (see Appendix 1). * Don't stack the trays one above the other or put them in direct contact with Code Description Proposed solution...
Page 11
10.6 Replace fuse 6.6 After the samples are loaded, close and lock the door by turning the door handle clockwise until it stops. 1). Switch off the power. Caution: You must turn the door handle to the maximum position, otherwise the machine will alarm and an error message will 2).
10.5 Replacement of the door seal 7. Draining the water tanks 10.5.1 Fully open the door. 10.5.2 Remove the door seal carefully by pulling it by hand away from the door 10.5.3 Clean the new door seal carefully with a soft cloth saturated with distilled water.
8 Advanced adjustment settings 10.3 Cleaning the inner-chamber, weekly Remove the trays and tray rack from the inner-chamber 8.1 Adjustment of sterilization time and drying time - Wipe the inner chamber with a soft cloth saturated in distilled water 8.1.1 Hold the UNWRAPPED button for 30seconds. The machine will beep - Apply the same procedure to the trays and tray rack.
9. Printer (Optional) 10 Maintenance 7. Maintenance 9.1 If you have purchase the BioClave printer device, you can connect the printer cable to the back of the sterilizer. 9.2 Connect the printer power cord to the socket with the proper electrical...