9. Printer (Optional)

9.1 If you have purchase the BioClave printer device, you can connect the
printer cable to the back of the sterilizer.
9.2 Connect the printer power cord to the socket with the proper electrical
specifications.
9.3 Power on the sterilizer and press the "sel" button.
9.4 After the sterilization cycle has completed, the printer will begin printing
the details of the sterilization cycle (see below).
Example of the print report
¨¨¨¨¨¨¨¨¨¨¨¨
Program:
wrapped
Temperature:
134
Pressure:
39 PSI
Vacuum Num.: 3
Dry Time:
15min
Ster Time:
12min
Cycle No:
00009
¨¨¨¨¨¨¨¨¨¨¨¨¨
MAX. Temp:
135.1
MIN. Temp:
134.1
¡
AVG. Temp:
134.6
MAX. Pres.:
41PSI
MIN. Pres.:
38PSI
AVG. Pres.:
39 PSI
¨¨¨¨¨¨¨¨¨¨¨¨¨
Start Time:
22:38
End Time:
23:17
Date:
2008-07-19
¨¨¨¨¨¨¨¨¨¨¨¨
11
8.1 Cleaning the door seal
8.2 Cleaning the clean water tank
Instructions manual
10 Maintenance
7. Maintenance
Frequency
Clean the door seal
Daily
Wipe dry the inner chamber
Clean the clean water reservoir
Weekly
Clean the inner-chamber
Annually
Replace the door seal
Clean the door seal daily with a softy cloth saturated in
distilled water
Clean the clean water tank every
week with medical disinfectant .