APPENDIX 2
DIAGRAMS OF THE STERILIZATION PROGRAMMES
PROGRAM
134
30
4
UNWRAPPED
121
16
15
134
30
6
WRAPPED
121
16
15
134
30
12
LIQUID
121
16
30
<100
0
15
DRY ONLY
The max. temperature of the 134 C sterilization cycle is 136C
The max. temperature of the 121 C sterilization cycle is 123C
1 9
TYPE
12~30
5.00
U n w r a p p e d s o l i d
m a t e r i a l
30~45
5.00
Unwrapped hollow
30~40
5.00
material
Single-wrapped solid
35~50
4.00
material
30~65
Wrapped liquid
0.50
Wrapped liquid
0.50
45~65
Instructions manual

4. Installation

*
Ensure that the sterilizer is installed with 2.5in. (10cm) ventilation space
on all sides of the sterilizer, and 5 in. (20cm) on top side. The clearance
required to open the door is 15.5in. (40cm).
*
The sterilizer should be placed on a level worktable.
*
Do not cover or block the door, ventilation or radiation openings on the
sterilizer.
Do not install the sterilizer near a sink or in a location where it is likely to
*
be splashed.
*
Do not install the sterilizer nearby a heat source.
Above dimensions are shown in mm.
Instructions manual
4