2.
PREPARE FOR TREATMENT
2.1 Place the device on a flat and sturdy surface in close proximity to where the
patient will be resting.
2.2 Attach the power cord to the back of the pump and then plug it into a safe 120 VAC,
60 Hz outlet. A grounded outlet is not required.
2.3 Note the LATCH CONNECTORS, numbered 1—4 and 5—8, located at the end of the
tubing on your garment (See Figure 2). With one hand squeeze the LATCH CON-
NECTOR with numbers facing up and push into the Receptor Ports (7) with matching
numbers located on the front panel of the device. You should hear a click when fully
engaged.
2.4 If two garments are used, remove the Blocker Bars (see Figure 3) and attach the
LATCH CONNECTORS from the second garment to the auxiliary ports (8).
2.5 Putting the garment on
LEG GARMENTS, unzip the garment gently all the way down to the bottom and stop
(zipper does not separate). Place the foot at the bottom end of the garment and pull
on garments straps to help guide the garment up and onto your leg. Once garment
is in place pull up the zipper while supporting the garment to wrap around the leg.
ARM GARMENTS, slide the arm through the internal cavity of the garment.
Note: Do not wear the garments directly over skin.
underneath garments for hygienic reasons and to avoid irritation.
recommended that light bandages, clean hosiery or stockinettes be worn
under garments. Clothing should be free of zippers, buttons or other items that
could rub and chafe the skin or damage the garment.
3.
Turning Pump On
3.1 Seat yourself comfortably in a reclined position, within easy reach of the pump. Legs
should be elevated (see photo on front page).
3.2 Powering on your device - Press the "Power On/Off" (1) button to turn your pump on
and off.
3.3 When first turned on, the LED Display (5) will show the total hours the pump has
been used. After 5 seconds, the LED display shows which mode your pump is set to,
after which the pump will start to blow air. The air pressure will be the same as the
previous treatment. Factory default pressure is 50 mmHg in the chamber #1, de-
creasing
to
pressure, see Section 5.
3.4 As the pump starts to run through its cycle, the number that shows in the LED
Display (5) is the countdown timer. The SC-3008-DL includes a automatic shutoff
timer. To run the pump in continuous (untimed) mode, see Section 6.
3.5 White operating in the Timed Shutoff mode, it is possible to view the set pressure by
holding down the "Mode" button. If you would like to toggle back to the countdown
timer, simple press the "Mode" button again.
43
mmHg
in
Always wear light clothing
chamber
#8.
9
It is
To
change
the
Need help?
Do you have a question about the SC-3008-DL and is the answer not in the manual?