4.
END OF TREATMENT
The pump powers down one of two ways:
1) After your 60 minute treatment is complete
2) You press the "Power On/Off" (1) button while the pump is running.
4.1 When your treatment is over, wait one minute before taking off the garment.
4.2 Squeeze the LATCH CONNECTOR and pull outward to remove garment from pump.
4.3 Continue to assist in the evacuation of air from garment, working from top to
bottom.
4.4 Once the garment feels loose enough, you can unzip the garment all the way to
bottom and remove.
Repeat steps 4.2 through 4.4 if two garments are used!
NOTE: In case of a power failure your device will automatically shut-down.
device does not have a battery backup and does not have timer memory. Once
your power is restored you will have to turn your device back on and
resume your treatment. A new 60 minute treatment will start. You can then end
the treatment manually by turning the power off.
GUIDELINES FOR TREATMENT
A physician is required to prescribe these settings, but general guidelines
are listed below:
All compression settings should be discussed with the physician. It is ultimately his / her
responsibility to prescribe the setting and it should be written on the
prescription upon referral.
physician is important when setting pressures.
60mmHg works well for most patients. However, a different pressure might be
prescribed for your personal needs.
Presence of fibrotic tissue may require as much as 80mmHg in order to soften the fi-
brotic tissue and achieve reduction. Once the tissue is soft, the compression can be re-
adjusted to 60mmHg.
Patients with a history of Congestive Heart Failure (CHF), which is controlled with medi-
cation, should never be in a flat position while pumping.
reclined position with elevated legs during treatment.
duration may be divided into twice a day 30 minutes per treatment.
Patients with a history of Deep Vein Thrombosis with or without a filter may
require less compression. These patients will generally tolerate 40mmHG. These
patients with a filter may need to divide their treatment into twice a day, 30
minutes per treatment.
Doppler study from the physician for their records.
To change the pressures in the individual garment chambers, See Section 5.
Every patient is unique and communication with the
It is suggested that the provider obtain a Negative
10
They should be in a
Their treatment regimen
The