Volts Verification - Keithley 617 Instruction Manual

Programmable electrometer
Hide thumbs Also See for 617:
Table of Contents

Advertisement

5.5.4
Volts
Verification
NOTE
Current
and
charge
verification
must
be per-
formed
before volts verification.
Verify the volts function
as follows:
1. Enable zero check and select the volts functions
with the
associated
front panel buttons.
2. Select the 200mV range and enable zero correct.
Check to
see that the display shows ooO.00 fl
count. If not, enable
zero correct.
3. Connect
the signal source to the instrument,
as shown
in
Figure 5-5. Set the calibrator
output to 190mV, as indicated
in Table 5-3.
4. Disable zero check and verify that the reading is within the
limits listed in Table 5-3.
Table
53.
Limits
for Volts
Verification
Applied
Allowable
Reading
Range
Calibrator
Input
I180
-28W
200mV
190.000mV
189.87 to 190.13mV
2 v
1.90000
v
1 .I?990 to 1.9010 v
20
v
19.0000
v
18.990 to 19.010 V
200
v
190.000 v
189.86 to 190.14 V
5. Repeat the procedure
for the ZV, ZOV, and 200V ranges by
applying
the respective
inputs listed in Table 5-3. Check to
see that the reading for each range is within the limits listed
in the table.
6. Repeat the procedure
for each of the ranges with negative
NOTE? LEAVE GREEN DISCONNECTED
SHORTING LINK REMOVED
LINK REMOVED
\
MODEL 617
/
Figure
5-5.
Connections
for Volts
Verification
5-5

Advertisement