Summary of Contents for DJO Global Vectra Genisys

  • Page 1 ® Leading Therapy In a New Direction User Manual Model 2784 Vectra ® Genisys Laser ISO 13485 Certified...
  • Page 2: Table Of Contents

    TABLE OF CONTENTS Vectra ® Genisys Laser FOREWORD ......... . 12 SETUP .
  • Page 3 TABLE OF CONTENTS TABLE OF CONTENTS Vectra ® Genisys Ultrasound Vectra ® Genisys Laser TREATMENT TIPS ........60 ACCESSORIES .
  • Page 4: Product Description

    FOREWORD TABLE OF CONTENTS Intelect ® Laser Module Vectra ® Genisys Laser This manual has been written for the operators of the Vectra ® Genisys Laser. It contains general instructions for operation, precautionary instructions, and maintenance recommendations. In order to obtain maximum life and efficiency from your Vectra ® Genisys Laser, and to assist in the proper operation of the unit, read and understand this manual thoroughly.
  • Page 5: Foreword

    FOREWORD Vectra ® Genisys Laser The following features are available on the Vectra ® Genisys Laser: • Clinical Portable Battery Powered Option The Vectra ® Genisys Laser is a truly portable laser unit that does not confine you to a wall socket to operate. •...
  • Page 6: Precautionary Instructions

    ABOUT LASER LIGHT THERAPY TABLE OF CONTENTS Intelect ® Laser Module Vectra ® Genisys Laser CAUTION Precautionary Instructions The precautionary instructions found in this section and • Read, understand, and practice the precautionary and operating throughout this manual are indicated by specific symbols. Understand these instructions.
  • Page 7: Cautions

    ABOUT LASER LIGHT THERAPY Vectra ® Genisys Laser CAUTION Failure to use and maintain the Vectra ® Genisys Laser and its accessories in accordance with the instructions outlined in this manual will invalidate • your warranty. • DO NOT remove the cover. This may cause unit damage, malfunction, electrical shock, fire, or personal injury. There are no user-serviceable parts inside the unit.
  • Page 8 ABOUT LASER LIGHT THERAPY TABLE OF CONTENTS Intelect ® Laser Module Vectra ® Genisys Laser WARNING • Use of controls or adjustments or performance of procedures other than those specified herein may result in hazardous exposure to laser light energy. •...
  • Page 9: Dangers

    ABOUT LASER LIGHT THERAPY Vectra ® Genisys Laser DANGER • Equipment not suitable for use in the presence of a flammable anesthetic mixture with air, oxygen, or nitrous oxide. • The solvents of adhesives and flammable solutions used for cleaning and disinfecting should be allowed to evaporate before the unit is used.
  • Page 10: Overview Of Laser Light Therapy

    ABOUT LASER LIGHT THERAPY TABLE OF CONTENTS Intelect ® Laser Module Vectra ® Genisys Laser OVERVIEW OF LASER LIGHT THERAPY In short, the light energy absorbed into the patient’s tissue triggers biological changes at a cellular level to provide topical heating for the temporary increase in local blood circulation;...
  • Page 11: Overview-Common Terms

    ABOUT LASER LIGHT THERAPY Vectra ® Genisys Laser OVERVIEW OF LASER LIGHT THERAPY Continuous Mode – The output of the laser light is not interrupted during the treatment time. Dosage – A measure of the intensity of the laser light energy over the treatment area. The unit of measure is Joules or Joules/cm Energy –...
  • Page 12 ABOUT LASER LIGHT THERAPY TABLE OF CONTENTS Intelect ® Laser Module Vectra ® Genisys Laser OVERVIEW OF LASER LIGHT THERAPY - COMMON TERMS (continued) Wavelength – Measured in nanometers (nm), wavelength is the key component in obtaining effective therapy as different wavelengths bring about different physiological effects.
  • Page 13: Indications

    ABOUT LASER LIGHT THERAPY Vectra ® Genisys Laser Indications for Laser Light Contraindications The Vectra ® Genisys Laser is indicated to provide topical The Vectra ® Genisys Laser should NOT be used: heating for the following: • where analgesia may mask progressive pathology, and where the •...
  • Page 14: Additional Precautions

    ABOUT LASER LIGHT THERAPY TABLE OF CONTENTS Intelect ® Laser Module Vectra ® Genisys Laser • over the thoracic area if the patient is using a cardiac DANGER pacemaker. Patients with an implanted neurostimulation device must not be treated • over or applied to the eye. with or be in close proximity to any shortwave diathermy, microwave diathermy, therapeutic laser diathermy or laser diathermy anywhere on •...
  • Page 15: Factors That Affect Treatment

    ABOUT LASER LIGHT THERAPY Vectra ® Genisys Laser discomfort. Choose a lower dosage to reduce output or select a pulsed duty cycle to decrease patient discomfort. Factors that Affect Treatment The following factors may affect laser light treatment: • Color of skin (light or dark) •...
  • Page 16 NOMENCLATURE TABLE OF CONTENTS Intelect ® Laser Module Vectra ® Genisys Laser UNIT LCD Intensity/ Contrast Dial Power On/Off Pause/Resume Laser Head TIME Up Arrow Clinical Resources LED Indicator (Output Power) STOP Frequency/Back Applicator (see page Down Arrow DOSAGE PAUSE START Intensity Display/Enter...
  • Page 17: Nomenclature

    NOMENCLATURE Vectra ® Genisys Laser APPLICATOR LED Indicator (Output Power) Lens Laser Diode Laser Head Pause/Resume Button Laser Head LED’s/SLD’s...
  • Page 18 NOMENCLATURE TABLE OF CONTENTS Intelect ® Laser Module Vectra ® Genisys Laser Power On/Off The Power On/Off button controls the flow of electricity to the unit. The LCD (Liquid Crystal Display) allows the user to view and monitor the information displayed during laser light therapy. The following information is displayed on the LCD: •...
  • Page 19 NOMENCLATURE Vectra ® Genisys Laser Frequency/Back Use this button to return to the previous window and toggle between 12 preset frequencies. Down Arrow When the window displays a list of options, press the Down Arrow button to scroll down the list. PAUSE Use this button to pause the treatment session.
  • Page 20 NOMENCLATURE TABLE OF CONTENTS Vectra ® Genisys Ultrasound Vectra ® Genisys Laser Applicator Connection This port serves as the connection point between the unit and the applicator. START Select Start to begin a treatment session. Intensity Display/Enter J/cm 2 Select this button to change display from J/cm to Joules.
  • Page 21 NOMENCLATURE Vectra ® Genisys Laser Charge Indicator This symbol displays when the unit is connected to mains power and the battery pack is charging. NOTE: During battery operation, if the unit is left on, but not active, for more than five minutes, it will power off to conserve battery power.
  • Page 22 NOMENCLATURE TABLE OF CONTENTS Vectra ® Genisys Ultrasound Vectra ® Genisys Laser LED’s/SLD’s These Light Emitting Diodes generate different wavelengths of light that allow the user to treat topical or surface symptoms. Laser Diode This mechanism generates different wavelengths of light that allow the user to treat various, deeper penetrating symptoms. LED Indicator (Output Power) This orange light illuminates when laser light energy is being distributed by the applicator.
  • Page 23: Specifications

    SPECIFICATIONS Vectra ® Genisys Laser UNIT SPECIFICATIONS DIMENSIONS Complies with 21CFR 1040.10 & 1040.11 IEC/UL/EN 60601-1, 60601-1-2, and 60601-2-22 Height (with base)............6.4 in (16.3 cm) IEC 60825-1:2001, CAN/CSA C22.2 No. 601.1-M90 w/A2 Width (with applicator) ..........11.3 in (28.8 cm) Each unit is shipped with repositional base, laser protective Width (without applicator) ........
  • Page 24: Laser Light Technical Specifications

    SPECIFICATIONS TABLE OF CONTENTS Intelect ® Laser Module Vectra ® Genisys Laser LASER TECHNICAL SPECIFICATIONS Duty Cycles The markings on the unit are assurance of its conformity to the highest applicable standards of medical Pulsed ......................90% equipment safety and electromagnetic compatibility. One Continuous ....................
  • Page 25: Laser Applicator Specifications

    SPECIFICATIONS Vectra ® Genisys Laser LASER APPLICATOR SPECIFICATIONS For all single diode and cluster laser and LED applicators, the expected increase in the measured quantities after manufacture added to the values measured at the time of manufacture is ±20%. The new software incorporates a cooling function that forces the user to cool the laser clusters prior to the next treatment. The software will calculate the cooling time needed when treatment times exceed 3 minutes per application.
  • Page 26: Laser Applicator Specifications

    SPECIFICATIONS TABLE OF CONTENTS Intelect ® Laser Module Vectra ® Genisys Laser LASER APPLICATOR SPECIFICATIONS continued Cluster Diode Applicators Applicator Output Power Contact Diode Diode Specifications Nominal Ocular Hazard Divergance Divergance Spot Size Power Density Area Type Distance (mW) (W/cm (NOHD-in meters for (rad) (rad)
  • Page 27: Laser Protective Eyewear Specifications

    SPECIFICATIONS Vectra ® Genisys Laser LASER PROTECTIVE EYEWEAR SPECIFICATIONS OPTICAL DENSITY vs. WAVELENGTH 1000 1100 Wavelength Useful Range Optical Density 5+ ..............190-400 nm Optical Density 3+ ..............625-830 nm Optical Density 3+ ..............815-1050 nm Each unit is shipped with Laser Protective Eyewear. The eyewear is L3 rated and approved and EN207 compliant. 0413...
  • Page 28: Setup

    TABLE OF CONTENTS SETUP Vectra ® Genisys Ultrasound Vectra ® Genisys Laser INSTALLING THE LASER INTERLOCK DOOR INTERRUPT SWITCH The Laser Interlock is an optional safety device designed to interrupt Laser Light Therapy anytime the door to the the therapy room is opened.
  • Page 29: Mounting The Unit On The Wall

    SETUP Vectra ® Genisys Laser MOUNTING THE UNIT ON THE WALL The Vectra ® Genisys Laser can be operated while the unit is resting on a flat surface, or mounted on a wall. To mount the unit on a wall, do the following: Remove the repositional base from the bottom of the unit.
  • Page 30 SETUP TABLE OF CONTENTS Vectra ® Genisys Ultrasound Vectra ® Genisys Laser MOUNTING THE UNIT ON THE WALL continued Using a 9/64 (3.6 mm or 0.357 cm) drill bit, drill four holes you marked in the previous step. Press 4 appropriately sized sheetrock anchors into the wall so that the sheetrock anchor is flush with the wall.
  • Page 31 SETUP Vectra ® Genisys Laser MOUNTING THE UNIT ON THE WALL continued Replace the repositional base on the bottom of the unit. Line up the screw heads with the holes on the repositional base, and slide the unit down slightly until the screw heads are securely fastened to the repositional base.
  • Page 32: Installing The Battery Pack

    TABLE OF CONTENTS SETUP Vectra ® Genisys Ultrasound Vectra ® Genisys Laser INSTALLING THE BATTERY PACK The Vectra ® Genisys Laser accommodates both AC mains power and an optional battery pack. The pack contains 20 Nickel Metal Hydride (NiMH) drycell batteries. The unit can operate with the rechargeable power supply for approximately five hours of continuous use.
  • Page 33: Installing The Battery Pack

    SETUP Vectra ® Genisys Laser INSTALLING THE BATTERY PACK (continued) Connect the battery pack cable to the unit’s battery connector in the bottom of the battery recess. Put the battery pack into the unit, making sure to orient it as shown. Replace the battery access door and re-tighten the screw using the screwdriver.
  • Page 34: Setup

    TABLE OF CONTENTS SETUP Vectra ® Genisys Ultrasound Vectra ® Genisys Laser CHARGING & USING THE BATTERY PACK CHARGING THE BATTERY PACK The battery pack is automatically charged by the unit whenever there is mains power connected. Charging may be interrupted during operation of the unit by the control circuitry to limit total power consumption.
  • Page 35: Operation

    OPERATION Vectra ® Genisys Laser ENTERING AND CHANGING THE PIN To gain access to any part of the Vectra ® Genisys Laser unit, you must enter a Personal Identification Number (PIN). The unit is shipped with a default PIN that allows you initial access, but you may change the number anytime. To enter and change the PIN, do the following: Turn the system power “ON”...
  • Page 36 OPERATION TABLE OF CONTENTS Vectra ® Genisys Ultrasound Vectra ® Genisys Laser ENTERING AND CHANGING THE PIN (continued) Press the Clinical Resources button. The Clinical Library window displays. Using the Up Arrow and Down Arrow buttons, highlight Change PIN. Press the Display (Enter) button to select the highlighted selection. The Change PIN window displays.
  • Page 37 OPERATION Vectra ® Genisys Laser ENTERING AND CHANGING THE PIN (continued) Press the buttons that represent the numbers to which you want to change the PIN. Once you enter four numbers, a message displays to inform you of the new PIN. Press any key on the Operator Interface.
  • Page 38 TABLE OF CONTENTS OPERATION Intelect ® Laser Module Vectra ® Genisys Laser PREPARING THE PATIENT’S SKIN FOR LASER LIGHT THERAPY Before applying laser light therapy to the patient, you must first prepare the patient’s skin. By properly preparing the patient’s skin for laser light therapy, you will allow more laser light energy to reach the targeted areas and reduce the risk of skin irritation.
  • Page 39: Starting, Stopping, & Interrupting Therapy

    OPERATION Vectra ® Genisys Laser STARTING, STOPPING, AND INTERRUPTING THERAPY The LCD will provide continuous information during the treatments concerning dosage and elapsed time. Parameters are adjusted using Operator Interface buttons on the front of the unit. The laser light output can be stopped by pressing the PAUSE or STOP buttons located on the Operator Interface.
  • Page 40 OPERATION TABLE OF CONTENTS Vectra ® Genisys Ultrasound Vectra ® Genisys Laser STARTING, STOPPING, AND INTERRUPTING THERAPY continued Press the DOSAGE button and raise or lower the unit’s output using the up and down arrows. NOTE: The Dosage and the Treatment Time are directly correlated. Whenever Dosage is changed, the Treatment Time is automatically changed and reflected on the window.
  • Page 41: Starting, Stopping, & Interrupting Therapy

    OPERATION Vectra ® Genisys Laser STARTING, STOPPING, AND INTERRUPTING THERAPY continued • the timer pauses • the unit beeps once • the icon displays • the laser applicator stops emitting laser light energy • the orange light goes out To resume therapy, press the PAUSE button on the Operator Interface, the Pause/Resume button on the back of the laser applicator, or the START button.
  • Page 42: Using Clinical Indications

    TABLE OF CONTENTS OPERATION Vectra ® Genisys Ultrasound Vectra ® Genisys Laser USING CLINICAL INDICATIONS The indications contained in this section are to be used only as guidelines. Each patient should be individually assessed to determine the appropriateness of the parameter setting prior to use. To select an indication for a patient, do the following: Press the Clinical Resources button.
  • Page 43 OPERATION Vectra ® Genisys Laser USING CLINICAL INDICATIONS (continued) Using the Up Arrow and Down Arrow buttons, highlight the appropriate indication. The indications displayed depend upon the applicator used. Press the DISPLAY (Enter) button to accept the highlighted selection. If available, highlight the appropriate laser light tissue depth with the indication you selected using the Up Arrow and Down Arrow buttons.
  • Page 44 TABLE OF CONTENTS OPERATION Vectra ® Genisys Ultrasound Vectra ® Genisys Laser USING CLINICAL INDICATIONS (continued) Press the DISPLAY (Enter) button to accept the highlighted selection. You are returned to the main window with the settings from the indication you selected displayed.
  • Page 45: Using Clinical Indications

    OPERATION Vectra ® Genisys Laser USING CLINICAL INDICATIONS (continued) To begin therapy, continue with the instructions outlined in the section entitled “Preparing the Patient’s Skin for Laser light Therapy” on page 35. Then, proceed to step 4 page...
  • Page 46: Creating A User Protocol

    TABLE OF CONTENTS OPERATION Vectra ® Genisys Ultrasound Vectra ® Genisys Laser CREATING A USER PROTOCOL This is a library you create. You may store up to 10 protocols in the User Protocol Library. To create User Protocols, do the following: Make the desired parameter changes.
  • Page 47: Creating A User Protocol

    OPERATION Vectra ® Genisys Laser CREATING A USER PROTOCOL (continued) Use the Up Arrow and Down Arrow buttons to highlight any unused user protocol. If you select Unit Default Protocol, this will become the protocol displayed when the unit powers up. Press the DISPLAY (Enter) button to accept the highlighted selection and save your custom protocol.
  • Page 48: Restoring Factory Settings

    OPERATION TABLE OF CONTENTS Vectra ® Genisys Ultrasound Vectra ® Genisys Laser RESTORING FACTORY SETTINGS Certain default utility settings on the unit may be changed to suit your requirements. These settings consist of the unit’s language and power up settings. However, you may want to return the unit to its original settings. To restore the original power up default settings on the unit, do the following: On the main window, press the Clinical Resources button.
  • Page 49: Restoring Factory Settings

    OPERATION Vectra ® Genisys Laser RESTORING FACTORY SETTINGS continued Press the DISPLAY (Enter) button to accept the highlighted selection. The Restore Factory Settings Confirmation window displays. Press any button on the Operator Interface. The default power up settings are restored and you are returned to the Clinical Library window.
  • Page 50: Restoring Factory Protocols

    TABLE OF CONTENTS OPERATION Vectra ® Genisys Ultrasound Vectra ® Genisys Laser RESTORING FACTORY PROTOCOLS If necessary, you can choose to restore the user-defined protocols to the unit’s original parameters when it was shipped to you. To do this, do the following: Press the Clinical Resources button.
  • Page 51: Restoring Factory Protocols

    OPERATION Vectra ® Genisys Laser RESTORING FACTORY PROTOCOLS continued) Press the DISPLAY (Enter) button to accept the highlighted selection. The Restore Factory Protocols Confirmation window displays. Press any button on the Operator Interface. The user-defined protocols are erased and restored to the original parameters. You are returned to the Clinical Library window.
  • Page 52: Selecting A User-Defined Protocol

    OPERATION TABLE OF CONTENTS Vectra ® Genisys Ultrasound Vectra ® Genisys Laser SELECTING A USER-DEFINED PROTOCOL To select a predefined laser light therapy program, do the following: Press the Clinical Resources button. The Clinical Library window displays. Use the Up Arrow and Down Arrow buttons to highlight the Retrieve User Protocol option.
  • Page 53: Selecting A User-Defined Protocol

    OPERATION Vectra ® Genisys Laser SELECTING A USER-DEFINED PROTOCOL (continued) Use the Down Arrow button to highlight the appropriate protocol. As you highlight each protocol, a description of the protocol’s parameters displays to the right. Press the Display (Enter) button to select the highlighted protocol. The main window displays with the parameters of the protocol you selected.
  • Page 54: System Utilities

    OPERATION TABLE OF CONTENTS Vectra ® Genisys Ultrasound Vectra ® Genisys Laser SYSTEM UTILITIES Audible Tones Audible tones will be heard in the following conditions: • The laser is arming • Any button is pressed. • The rechargeable battery’s power is low (in which case the Low Battery icon will display). •...
  • Page 55 OPERATION Vectra ® Genisys Laser SYSTEM UTILITIES continued Make the desired changes. Press the Clinical Resources button. The Clinical Library window displays. Press the Up Arrow and Down Arrow buttons to highlight Save Protocol, and press the DISPLAY (Enter) button to accept the highlighted selection. The Save Protocol menu displays.
  • Page 56: Brightening Or Dimming The Lcd

    OPERATION TABLE OF CONTENTS Vectra ® Genisys Ultrasound Vectra ® Genisys Laser SYSTEM UTILITIES (continued) Press the DISPLAY (Enter) button to accept the highlighted selection. The User Default Protocol confirmation window displays. Press any key to confirm the settings. You are returned to the Clinical Library window. Brightening or Dimming the LCD To brighten or dim the LCD, turn the contrast control dial until the display contrast is optimal.
  • Page 57: Viewing Applicator Information

    OPERATION Vectra ® Genisys Laser SYSTEM UTILITIES (continued) Viewing Applicator Information Use this utility to verify applicator specifications. To do this, do the following: Press the Clinical Resources button. The Clinical Library window displays. Press Up Arrow and Down Arrow buttons to highlight the View Applicator Info option.
  • Page 58 OPERATION TABLE OF CONTENTS Vectra ® Genisys Ultrasound Vectra ® Genisys Laser SYSTEM UTILITIES (continued) Press the DISPLAY (Enter) button to accept the highlighted selection. The Laser Applicator Information window displays. Verify the information on the window and press the Back button to return to the Clinical Library window.
  • Page 59: Changing Languages

    OPERATION Vectra ® Genisys Laser SYSTEM UTILITIES (continued) Changing Languages You may change the language displayed by the Vectra ® Genisys Laser to either English or Spanish. To change the language displayed on the LCD, do the following: Press the Clinical Resources button. The Clinical Library window displays.
  • Page 60 TABLE OF CONTENTS OPERATION Vectra ® Genisys Ultrasound Vectra ® Genisys Laser SYSTEM UTILITIES continued) Press the DISPLAY (Enter) button to accept the highlighted selection. Press the Up Arrow and Down Arrow buttons to highlight the appropriate language. Press the DISPLAY (Enter) button to accept the highlighted selection. Your unit now displays the language you selected.
  • Page 61: Viewing Unit Version Information

    OPERATION Vectra ® Genisys Laser SYSTEM UTILITIES continued) Viewing Unit Version Information Use this utility to verify that the unit is using the latest software available. To do this, do the following: Press the Clinical Resources button. The Clinical Library window displays. Use the Up Arrow and Down Arrow buttons to highlight the View Unit Version Info option.
  • Page 62: System Utilities

    OPERATION TABLE OF CONTENTS Vectra ® Genisys Ultrasound Vectra ® Genisys Laser SYSTEM UTILITIES (continued) Press the Display button to accept the highlighted selection. The Unit Version Information window displays. Press any key to return to the Clinical Library window.
  • Page 63: Treatment Tips

    OPERATION Vectra ® Genisys Laser TREATMENT TIPS Contact To obtain the most effective results, the applicator should be in contact with the patient’s skin. Applicator Position Due to the characteristics of laser light, the angle at which the light enters the patient’s skin is very important. Therefore, the applicator lens should always be parallel to the treatment area.
  • Page 64: Accessories

    TABLE OF CONTENTS ACCESSORIES Intelect ® Laser Module Vectra ® Genisys Laser Standard Accessories Optional Accessories Ref. Description Ref. Description 28066 User Manual (CD-ROM) 27805 Single Diode 820 nm Laser 300 mW Applicator [26 AWG, 68.89 in (174.143 cm), Shielded] 27525 Laser Protective Eyewear 27840...
  • Page 65: Troubleshooting

    TROUBLESHOOTING Vectra ® Genisys Laser ERRORS Troubleshooting the Display If you press the Power On/Off button, and the LCD remains blank longer than a few seconds, the contrast may require adjusting. To adjust it, turn the contrast control dial clockwise until the display contrast is optimal.
  • Page 66: Maintaining The Unit

    Intelect ® Laser Module Vectra ® Genisys Laser Maintaining the Vectra Genisys Laser The following items should be checked at least monthly to ensure proper operation of this unit: • Power cord and plug: Check to make sure the cord is not frayed, kinked, and does not have torn or cut insulation.
  • Page 67: Service

    MAINTENANCE Vectra ® Genisys Laser Service The Vectra ® Genisys Laser applicators must be recalibrated annually. It is recommended that all Chattanooga Group laser products be returned to the factory or an authorized servicing dealer for repairs or recalibration. Recalibration is also recommended after the replacement or repair of any major component.
  • Page 68: Warranty

    WARRANTY TABLE OF CONTENTS Intelect ® Laser Module Vectra ® Genisys Laser Chattanooga Group ("Company") warrants that the Vectra ® Genisys Laser ("Product") is free of defects in material and workmanship. This warranty shall remain in effect for two years (24 months) from the date of original consumer purchase. If this Product fails to function during the two year warranty period due to a defect in material or workmanship, Company or the selling dealer will repair or replace this Product without charge within a period of thirty (30) days from the date on which the Product is returned to the Company or the dealer.
  • Page 69 Leading Therapy In a New Direction ISO 13485 Certified 4717 Adams Road P.O. Box 489 Hixson, TN 37343 U.S.A. (800) 592-7329 - U.S.A. (800) 361-3661 - Canada (800) 242-8329 chattgroup.com © 2006 Encore Medical, L.P. 28050B...