Specifications
General Specifications
Circuit
Display
Units of Measure
Input circuit
Data Hold
Sensor
Zero Adjust
Sample rate
Compatibility
Auto Power OFF
Data Output
Ambient conditions
Battery power
Power Consumption 6mA DC (approx.)
Dimensions
Weight
Accessories
Range Specifications
Range/Units (Max.)
± 2000 mbar
± 29 psi
Custom one-chip LSI microprocessor circuit
0.6" (16mm) 2000 count LCD with function indicators
2
mBAR, psi, kg/cm
, mm Hg, inches Hg, meter H
and Atmospheres
Differential inputs (P1 and P2)
Freezes displayed reading
Built-in piezoelectric sensors
Front panel push-button
0.8 sec. (approx.)
Air or non-corrosive and non-ionized gas
Meter turns off after approx. 10 minutes to conserve energy
RS-232 PC serial interface
o
o
32 to 122
F (0 to 50
C); < 80% RH
9V alkaline battery
7.1 x 2.8 x1.3" (180 x 72 x 32 mm)
0.76 lbs (345 g)
Quick-disconnect fittings (2), rubber jacket, and 9V battery
Resolution
1
0.02
3
O, inches H
2
Accuracy
Model 407910 - Version 2.0 - March 2005
O
2