GE Oil & Gas Dresser Series B3 Meter with ES3 Electronic Temperature Compensator Circular Pulse Output Connector Installation Supplement imagination at work...
ROOTS* ES3 Electronic Temperature Compensator (TC) Installation Procedures Use and Limitations 1. Meter Installation This document provides recommendations where there Refer to “Installation Supplement IS:B3” for Meter is no established company procedure or practice. installation procedures. WARNING WARNING If equipment is installed/serviced/maintained at This equipment is designed to operate at temperatures elevated heights, ensure proper safe site work practices between -40°...
Page 3
Table 1 - Scrolling sequence for ES3 screen display Displayed on Screen Represents Function Displays non-compensated volume which has COMPENSATED VOLUME Compensated Volume been corrected to standard conditions NON-COMPENSATED VOLUME Non-compensated Volume Displays actual non-compensated volume LINETEMP Line Temperature Displays live line temperature FIXED P Fixed Line Temperature Displays the line pressure as entered by the user...
Page 4
4. Wiring the Pulse Outputs Table 2 - Circular Connector Wiring See Table 2 and Figure 4 for output pulse wiring for a Output Name Connector Pin Pulse Type Circular Connector. Pulse Output 1 (+) PO1 (+) Form A WARNING Pulse Output 1 (-) PO1 (-) Ensure properly licensed/trained professionals are used...
Page 6
After clicking “Test Pulse Outputs,” a screen appears as shown in Figure 8. Click “Yes” to proceed with the pulse output test. For further information, refer to the Dresser MeterWare Manual. GE Oil & Gas Figure 8- Send test Pulses screen in Dresser MeterWare software 16240 Port Northwest Drive, Suite 100 Houston, TX 77041 P: +1 832.590.2303 Inside US P: 800.521.1114...