If this should occur, re-zero the instrument as follows:
1. Remove any attached accessories
and select Moisture Search Mode
(see section 2.1 for instructions).
MOISTURE & REL is displayed.
2. With one hand, hold the instrument
at least 12 in. (0.3 m) from all
objects (see the figure to the right).
3. With your other hand,
simultaneously press the
buttons. Hold them down until the
display shows nul, then Air. Quickly
release the buttons and continue to
hold the instrument as shown, well
clear of your body and your other
hand. Keep the instrument in free
air (i.e. at least 12 in. (0.3 m) from
all objects) while the display is
showing Air.
4. After about 10 seconds, the display should return to normal and the
instrument should read zero when held in free air.
You may now take normal Search Mode readings with the MMS.
2.1.1.1 Troubleshooting the Zeroing Function - The instrument
changes mode when zeroing Is attempted
If the instrument changes mode when you attempt to zero it, you have most
likely not pressed the two buttons at exactly the same time. Return to Search
Mode and attempt to press and hold the two buttons simultaneously.
2.1.1.2 Troubleshooting the Zeroing Function - The instrument
shows Air, but then displays Err or E30 and shuts OFF
If this situation occurs, repeat the zeroing procedure. If this does not solve the
problem, it is likely that the temperature adjustment is outside the permitted
range. If possible, allow the unit to return to normal room temperature (about
20°C/68°F). If the unit will not re-zero at this temperature, the unit is likely
faulty and requires repair/service.