Edit User Data; Add Apparatus Models - Honeywell Posi3 USB Setup Manual

Computerized performance tester for air-supplied breathing apparatus
Hide thumbs Also See for Posi3 USB:
Table of Contents

Advertisement

6.3.3

Edit User Data

To edit an existing user's information, enter
the User List as described above and click on
the up arrow.
Once the user's information has been
changed, click on the Check button to save
the information.
6.4

Add Apparatus Models

Note: The Model Configuration Wizard is only
available in Standard Version Posi3 USB
software. Brand-specific versions of software
come with pre-installed models and do not
allow end users to configure models.
All SCBA are made up of similar components
including a facepiece, a first stage regulator, a
second stage regulator, alarms, gauges and
an air cylinder. Although every SCBA is
unique, there are general similarities from
model to model that Honeywell Analytics uses
to set the appropriate baseline values for
testing these components.
The Model Configuration Wizard allows the
user to select components to create specific
models from the available of SCBA, Airline
Apparatus, and Facepieces that are used in
testing.
Once the type of apparatus is selected, the
Model Configuration Wizard will automatically
run through the appropriate setup procedure.
The user will need to configure the model so
that it matches the physical specifications of
the apparatus.
The testing procedure is automatically defined
for the apparatus model that is selected at the
time of testing. That apparatus is configured
in the software by this procedure.
The user entering apparatus information must
have level 4 software access or the Add New
Model and/or Add New Serial Number options
may be unavailable. See section 6.3.1 for a
description of user software permissions.
1. To enter a new apparatus, first select
Setup / Apparatus / Add New Model.
2. The Choose Model Screen will be shown
and will show a list of all models that are
available for testing. If this is the initial
software launch, the fields will be blank.
3. Click on New to enter the Model
Configurator.
4. Enter the Manufacturer Name and the
Model Name. Then select the units
(Metric, Imperial or Pascal), select the
pressure of the cylinder (2216 PSI, 3000
PSI or 4500 PSI) and select the type of
apparatus (SCBA, Airline Apparatus,
SCBA/Airline or Facepiece).