Re-Computing The Ci-203 Parameters - BCID Bio-Science CI-203 Operation Manual

Handheld laser leaf area meter
Table of Contents

Advertisement

Re-computing the CI-203 Parameters

This procedure should be carried out whenever the CI-203's threshold settings are changed or if the
instrument isn't measuring accurately. Re-computing the parameters will recalculate the effects of
any changes made on the other parameters.
1. From the "Measure" menu, press the LEFT and RIGHT arrow keys at the same time.
2. Hold the buttons until the motor begins to spin. The motor should spin for about 4 seconds.
3. Do this twice by pressing the LEFT and RIGHT arrow keys simultaneously again and making
sure that the motor spins up.
4. Take a test measurement and try to determine if the Length or Width readings need
adjustment. Are the readings too high or too low?
1554 NE 3
Ave, Camas, WA 98607, USA
rd
Phone (360) 833-8835 Fax (360) 833-1914
sales@cid-inc.com http://www.cid-inc.com
CI-203 Instruction Manual rev. 8/9/2016
27