The following error messages should rarely appear in normal operation. If they do appear, see
Appendix III and follow the steps to restore the setup data.
MESSAGE:
Bad NVRAM Checksum <start>
Leaf Not Cal'ed <start>
Root Not Cal'ed <start>
Conveyor Not Cal'ed <start>
Bad EPROM Checksum
1554 NE 3
Ave, Camas, WA 98607, USA
rd
Phone (360) 833-8835 Fax (360) 833-1914
MEANING:
Instrument setup data has been lost. Pressing
the START key will restore default values and
display the message "NVRAM Defaults Saved".
Measurement will not be accurate until re-
calibrated (See Appendix III).
Conveyor, Leaf or Root mode selected when
the power turned on has not been calibrated.
Pressing the START key will dispatch the
message and allow the instrument to operate,
but measurements will not be accurate until
calibrated. This message will appear every
time the power is turned on, until the
instrument is calibrated.
EPROM containing instrument firmware has
bad checksum, indicating it has failed. The
instrument will not operate with a bad
EPROM.