Testing The Ac Voltage Function - Fluke 80 SERIES III Service Information

Hide thumbs Also See for 80 SERIES III:
Table of Contents

Advertisement

80 Series III
Service Information
Rotary Switch Position

Testing the AC Voltage Function

To verify accuracy in the ac voltage ranges, do the following:
1. Connect the Calibrator to the
2. Set the Calibrator for the voltage and frequency called for in step 1 of Table 4.
3. Compare the reading on the Meter display with the display reading for your 80
4. Repeat this procedure for steps 2 through 17 of Table 4.
Step
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
16
Display (±12)
OFF
V ac
V dc
mV dc
Series III model (83, 85, or 87) shown in Table 4.
If the display reading falls outside of the range shown in Table 4, the Meter does not
meet specification.
Input
Voltage
350.0 mV
350.0 mV
350.0 mV
350.0 mV
3.500 V
3.500 V
3.500 V
3.500 V
35.00 V
35.00 V
35.00 V
35.00 V
350.0 V
350.0 V
350.0 V
900V
900 V
Table 3. Rotary Switch Test
Rotary Switch Position
N/A
Continuity/Ohms
0
Diode test
-32
mA/A
µA
-64
z
and
COM
Table 4. AC Voltage Test
Frequency
83 Series III
60 Hz
347.8 to 352.2
1 kHz
346.1 to 353.9
5 kHz
342.6 to 357.4
20 kHz
60 Hz
3.480 to 3.520
1 kHz
3.461 to 3.539
5 kHz
3.426 to 3.574
20 kHz
60 Hz
34.80 to 35.20
1 kHz
34.61 to 35.39
5 kHz
34.26 to 35.74
20 kHz
60 Hz
348.0 to 352.0
1 kHz
346.1 to 353.9
2.5 kHz
342.6 to 357.4
60 Hz
1 kHz
Display (±12)
inputs on the Meter.
Display Reading
85/87 Series III
347.1 to 352.9
346.1 to 353.9
342.6 to 357.4
NA
341.0 to 359.0
3.473 to 3.527
3.461 to 3.539
3.426 to 3.574
NA
3.410 to 3.590
34.73 to 35.27
34.61 to 35.39
34.26 to 35.74
NA
34.10 to 35.90
347.3 to 352.7
346.1 to 353.9
342.6 to 357.4
893 to 907
892 to 908
887 to 913
887 to 913
-96
-128
-160
-192