Testing The Milliamp (Ma) Function; Testing The Microamp (Μa) Function - Fluke 80 SERIES III Service Information

Hide thumbs Also See for 80 SERIES III:
Table of Contents

Advertisement

If you use a Fluke 5100 Series Calibrator, activate the 50 Ω divider
override.
3. Apply 3.000 V.
The Meter display should read 3.000 V dc +/- 0.061 V dc.

Testing the Milliamp (mA) Function

To verify the accuracy of AC and DC current measurement functions, do the following:
1. Connect the Calibrator to the
2. Turn the rotary switch to
The Meter enters the DC measurement function
3. Apply the inputs in steps 1 and 2 in Table 11.
4. For each input, compare the readings on the Meter display to the display readings for
your Model of Meter.
5. Press the blue button on the Meter to toggle to AC measurement function.
6. Apply the inputs in steps 3 through 6 in Table 11.
7. For each input, compare the readings on the Meter display to the display readings for
your Model of Meter.
Step
DC Current
1
35.00 mA
2
350.0 mA
AC Current
3
35.00 mA
4
35.00 mA
5
350.0 mA
6
350.0 mA
Testing the Microamp (µA) Function
To verify the accuracy of the microamp (
1. Connect the Calibrator to the
2. Turn the rotary switch to
The Meter enters the DC measurement function.
3. Apply the inputs in steps 1 and 2 of Table 12.
4. For each input, compare the readings on the Meter display to the display readings for
your Model of Meter.
5. Press the blue button on the Meter to toggle to the AC measurement function.
6. Apply the inputs in steps 3 and 6 of Table 12.
Note
mA µA
mA/AC.
Table 11. mA Tests
83 III
34.82 to 35.18
348.4 to 351.6
Frequency
83 III
60 Hz
34.56 to 35.44
1.0 kHz
34.56 to 35.44
60 Hz
345.6 to 354.4
1.0 kHz
345.6 to 354.4
mA µA
µAC.
and
COM
inputs on the Meter.
Display Reading
85 III
34.89 to 35.11
349.1 to 350.9
85 III
34.63 to 35.37
34.63 to 35.37
346.3 to 353.7
346.3 to 353.7
µA
) measurement function, do the following:
and
COM
inputs on the Meter.
Multimeters
Performance Tests
87 III
34.89 to 35.11
349.1 to 350.9
87 III
34.63 to 35.37
34.63 to 35.37
346.3 to 353.7
346.3 to 353.7