QUICKSTART

Your TV2 can be installed and operating in just a few minutes by following these simple steps.
1. Connect the USB cable to a PC or wall plug and the System Parameter display will appear.
2. Touch 'Up' or 'Down' to select C° or F° and touch 'Enter'.
3. Touch 'Up' to select the date format (MM/DD/YY, DD/MM/YY or YY/MM/DD) and then
'Enter'. The date format defaults to the US style (MM/DD/YY).
4. Set the day, month, year, hour and minute in the appropriate field by scrolling through the
possible values with 'Up' or 'Down' after moving to the appropriate field with 'Enter'.
(Note: It is possible to move to another field with the 'Right' button but your entries will
not be recorded).
5. Touch 'Escape'. You will be asked to verify that you want to erase all data. Use 'Right' to
highlight YES and touch 'Enter'.
6. Touch 'Escape' to return to the previous menu.
The TV2 Room Pressure Monitor is now operational