3.5 Power Supply Connections
Warning.
Disconnect the supply from any cables being terminated on the transmitter.
Electrical installation and earthing (grounding) must be in accordance with relevant national and
local standards.
Note.
Power supply connections / earthing arrangements are identical for cathodically-protected remote
transmitter systems. For cathodically-protected integral transmitter systems, follow cathodic
installation practices.
AquaMaster 3 has 3 power supply options:
–
Mains power – see Section 3.5.1
–
Battery power – see Section 3.5.2, page 29
–
Renewable energy – see Section 3.5.3, page 30
3.5.1 Mains Power Supply
Note. Before making connections, check the Data label to confirm power supply requirements.
Mains power requirements:
110 to 240 V AC, 50 / 60 Hz @ <3 VA
Cable length 3 m (9.8 ft.)
Protected by a fused isolator, rating – mains, anti-surge 3 A.
Make connections as shown in Fig. 3.17.
Color
Green / Yellow
Brown
Blue
110 to 240 V AC
50 / 60 Hz
Fig. 3.17 Connecting a Mains Power Supply
28
Connection
Earth
L1 / L
L2 / N
Isolator /
Fuse
Rating – mains, anti-surge, 3 A