Ohms Measurement
Range
Resolution
400.0 Ω
0.1 Ω
4.000 kΩ
0.001 kΩ
40.00 kΩ
0.01 kΩ
400.0 kΩ
0.1 kΩ
4.000 MΩ
0.001 MΩ
40.00 MΩ
0.01 MΩ
Overload protection: 1000 V
Open circuit voltage: <3.9 V
42
info@Fluke-Direct.com
Accuracy, ±(% of Reading + Counts)
Measurement Current
220 μA
60 μA
6.0 μA
600 nA
220 nA
22 nA
Fluke-Direct
0.2 % + 2
0.2 % + 1
0.2 % + 1
0.2 % + 1
0.35 % + 3
2.5 % + 3