9. Press
Calibrate.
10. Press
A reverse font (light background
with dark characters) will appear
to indicate that the reading can
be adjusted.
11. Press
to the concentration of the
standard, 1.00 in this example.
Note: The allowable adjustment
is ±25%.
12. Press
Two menu choices will be
offered, Set Calibration and
Factory Setting.
13. Press
Calibration and save the
calibration. Press
to scroll to and select Factory
Setting to revert to the factory
calibration. The meter will
momentarily display Storing...
and return to the Chlorine TT
Menu. The calibration has now
been saved and the meter can
be used for testing.