The user selects a Relay by scrolling up/down the list of available Relays until the
desired Relay is highlighted and then presses the (u) [Enter] button. The follow-
ing screen will appear:
The user can configure the Relay action from 9 different options:
No Action
Totalizer > Limit
High Flow Alarm
Low Flow Alarm
Range between H&L : (R) Range between High and Low Flow Alarm condition.
High Temp. Alarm
Low Temp. Alarm
High Pres. Alarm
Low Pres. Alarm
Exit
The user selects an Action by scrolling up/down the list of available options until
the desired option is highlighted and then presses the (u) [Enter] button.
Figure b-15, Relay Assignment Screen
Figure b-16, Relay #1 Action Settings
: (N) No assignment (relay is not assigned to any events).
: (T) Totalizer reached set limit volume.
: (H) High Flow Alarm condition.
: (L) Low Flow Alarm condition.
: (A) High Temperature Alarm condition.
: (B) Low Temperature Alarm condition.
: (C) High Pressure Alarm condition.
: (D) Low Pressure Alarm condition.