CP AL1 User Manual
Table 2-1:
Setup screen settings
20
The available settings are described below.
Setting
AUTO POWER OFF Defines the time the CP AL1 is switched off
AUTO LIGHT OFF
LCD CONTRAST
MULTIPLE SETUP
Note: To load the default settings after switching the CP AL1 off, press and hold
the ESC button, and then press the power on/off key for at least 2 sec.
Description
automatically after the last key press. The available
values are: 3 MIN, 10 MIN, 30 MIN, 60 MIN and
DISABLE. If DISABLE is selected, the CP AL1 is
turned on until it is switched off by pressing the power
on/off button.
Defines how long the backlight stays on after any key
press. The available values are: 3 SEC, 10 SEC,
60 SEC and DISABLE. If DISABLE is selected, the
backlight stay on until the CP AL1 is switched off.
Note: Lighting of the backlight shortens the battery life.
Adjusts the contrast of the display.
Note: Alternatively, you can adjust the contrast if a
measurement result is displayed by holding the ESC
button down and pressing the up/down arrow key.
Allows storing up to four individual settings.
To enable the MULTIPLE SETUP mode, set the
corresponding value to ENABLE. The next time the
CP AL1 is switched on, select the setting ID (1, 2, 3 or
4) on the start screen. After the CP AL1 is switched off,
the setting values are stored under this ID.