12
.
Measurement
Instructions
Perform
DC
ZERO,
if
necessary.
No
additional
zero procedure
is
needed
for ac
voltage.
Select desired modifiers
and
memory
operations.
2-83.
DC
VOLTAGE
(V DC)
2-84.
Use
the following
procedure
when
measuring
dc
volts:
1.
Select
the
VDC
function.
2.
The
8502A
will
go
to autoranging.
If
desired,
select
manual
ranging
and one
of
the
five available
ranges
(lOOOV, lOOV, lOV, IV,
or
100
mV).
3.
Perform
DC
Zero,
if
needed (100
mV
range
only).
4.
Select desired
modifiers
and
memory
operations.
5.
Connect
the
dc
voltage to the
HI and
LO SENSE
INPUT
terminals.
6.
A
dc
voltage reading
should
now
appear
in
the
display.
2-85.
DC
ZERO
2-86.
For dc
zeroing,
use
the
following
steps:
L
Select
the
V
DC
function
and
either
autoranging
or the
100
mV
manual
range.
2.
Place a
good
quality,
low
thermal
short across
the
H! and
LO
INPUT
terminals.
3.
Depress the
STORE
switch.
"?"
will
appear
in
the
display.
4.
Depress
the
VDC/OHMS ZERO
switch;
value
applied to
input terminals
will
now
be displayed
as
long
as
the
switch
is
held depressed.
5.
Release
of
the
switch
will
activate
V
DC/Ohms
Zero
mode,
applying
the value read
in
step
4
as
a
zero reference
for
subsequent
readings.
2-87.
AC
VOLTAGE
(V
AC)
2-88.
Use
the following
procedure
when
measuring
ac
volts:
1.
An
AC
Converter
module
must
be
installed
(Option
—01,
or
—09
A).
2.
Select the
VAC
function.
3.
The
8502A
will
go
to
autoranging;
manual
ranging
can
be
selected
if
necessary.
There
are
four
avail-
able ranges:
lOOOV, lOOV,
lOV,
IV.
5.
6
.
The
slow
filter
(FILTER
LED
illuminated)
must
be
selected for
full
accuracy
below 400 Hz.
7.
Connect
the
ac voltage to the
HI and
LO SENSE
INPUT
terminals.
8.
The
ac voltage
should be read
in
the
display.
2-89.
AC VOLTAGE ON
A DC
LEVEL(VDC
and
V
AC)
2-90.
Use
the following
procecure
when
measuring
ac
and dc
voltage;
1.
The
RMS
AC
Converter
module must
be
installed.
2.
Depress
both
V
DC
and
V
AC
awitches simultane-
ously.
3.
The
8502A
will
go
to autoranging;
manual
ranging
can be
selected
if
necessary.
There
are
four
avail-
able ranges:
lOOOV, lOOV,
lOV
or
IV.
4.
Select desired modifiers
and
memory
operations.
5.
Connect
the
unknown
voltage to the
HI
and
LO
SENSE
INPUT
terminals.
The
reading displayed
will
be the
rms
value
of
the
two
voltages
combined.
2
-
91
.
DC
CURRENT
(A
DC)
2-92.
Use
the following
procedure
when
measuring dc
current:
1.
The
Current Shunts
module
must
be
installed.
2.
Select the
A
DC
function.
3.
The
8502A
will
go
to autoranging;
manual
ranging
can
also
be
utilized
to select
one
of
five
available
ranges
(100 uA,
1
mA,
10
mA,
100
mA,
or lA).
4.
Select desired modifiers
and
memory
operations.
5.
Connect dc
current to
HI and
LO
INPUT
SOURCE
terminals.
2-93.
AC
CURRENT
(A
AC)
2-94.
Use
the following
procedure
when
measuring
ac
current:
1.
The
Current Shunts
module
must be
installed.
2-18

Advertisement