Technical Data - BRAND Titrette Manual

Class a precision digital burette
Hide thumbs Also See for Titrette:
Table of Contents

Advertisement

Technical data

Comparison of error limits
Titrette
Volume
Partial
A*
≤ ± %
ml
volume, ml
10
10
0,10
N E W !
5
0,20
1
1,00
25
25
0.07
12.5
0.14
2.5
0.70
50
50
0.06
25
0.12
5
0.60
* Error limits related to the nominal capacity (= maximum volume) indicated on the instrument, obtained when instrument and
distilled water are equilibrated at ambient temperature (20 °C/68 °F) and with smooth operation.
** Error limit: EL = A + 2CV, according to DIN EN ISO 8655-6 Annex B
(A = accuracy, CV = coefficient of variation, EL = error limit)
The titration volume is displayed in steps of 1 µl at instruments with 10 ml
and 25 ml size and in steps of 2 µl for 50 ml size instruments. For titration
volumes above 20 ml the display will automatically switch to steps of 10 µl.
Material and reagents
The instrument can be used for the following titration media
(maximum concentration 1 mol/l):
Acetic acid
Alcoholic potassium hydroxide
solution
Ammonium iron (II) sulfate solution
Ammonium thiocyanate solution
Barium chloride solution
Bromide bromate solution
Cerium (IV) sulfate solution
EDTA solution
Hydrochloric acid
Hydrochloric acid in Acetone
Iodide Iodate solution*
Iodine solution*
* Use light shield inspection window
When the instrument is properly handled, dispensed liquid will only come into contact with the
following chemically resistant materials: borosilicate glass, Al
platinum-iridium; PP (screw cap).
Limitations of use
Chlorinated and fluorinated hydrocarbons or
chemical combinations which form deposits
may make the piston difficult to move or may
cause jamming.
Compatibility of the instrument for a special
application (e.g., trace material analysis) must
be checked by the user. For additional infor-
mation, please contact the manufacturer.
The instrument is not autoclavable!
®
bottle-top burette
Bottle-top burettes acc.
to DIN EN ISO 8655-3
CV*
A*
≤ %
≤ ± %
µl
µl
µl
10
0,05
5
0,3
10
0,10
5
0,6
10
0,50
5
3
18
0.025
6
0.2
18
0.05
6
0.4
18
0.25
6
2
30
0.02
10
0.2
100
30
0.04
10
0.4
100
30
0.20
10
2
100
Note:
If you need an official certification which confirms the
error limits that are much stricter than those of DIN EN
ISO 8655-3, we recommend a calibration certificate
from an accredited calibration laboratory (e.g., the
DAkkS laboratory at BRAND).
Iron (II) sulfate solution
Nitric acid
Oxalic acid solution
Perchloric acid
Perchloric acid in glacial acetic
acid
Potassium bromate solution
Potassium bromide bromate
solution
Potassium dichromate solution
Potassium hydroxide solution
Potassium iodate solution
Potassium permanganate solution*
Operating limits
This instrument is designed for titrating liquids,
observing the following physical limits:
+15 °C to +40 °C (59 °F to 104 °F)
of instrument and reagent
Vapor pressure up to 500 mbar
Viscosity up to 500 mm
Altitude: maximum 3000 m above sea level
Relative humidity: 20% to 90%
Glass burettes Class A
acc. to DIN EN ISO 385
and ASTM
CV*
EL**
≤ %
µl
± µl
30
0,1
10
20
30
0,2
10
20
30
1
10
20
50
0.1
25
30
50
0.2
25
30
50
1
25
30
0.1
50
50
0.2
50
50
1
50
50
Potassium thiocyanate solution
Silver nitrate solution*
Sodium arsenite solution
Sodium carbonate solution
Sodium chloride solution
Sodium hydroxide solution
Sodium nitrite solution
Sodium thiosulfate solution
Sulfuric acid
Tetra-n-butylammonium
hydroxide solution
Triethanolamine in Acetone*
Zinc sulfate solution
O
, ETFE, PFA, FEP, PTFE,
2
3
/s
2

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents