Elcometer 456 Operating Instructions Manual page 6

Miniature probe
Hide thumbs Also See for 456:

Advertisement

When a new type of probe is fitted to the Elcometer 456 gauge the gauge factory reset feature should be
used to set the gauge to the calibration conditions set in the memory of the probe during its final test in the
factory. The following procedure will activate the factory reset:
Note: This procedure is the same for all variants of the Elcometer 456, Basic, Standard and Top.
1. Switch the gauge on by pressing the e key.
2. Select MENU from the READING SCREEN
3. Select RESET
4. Select FACTORY RESET
5. Select YES
4
Reading Screen for the Elcometer 456 Top
with Statistics Off.
Probe type
FM3 or NM3
is displayed
here