Download Print this page

Revolutionary Science Saniclave RS-SC-102 Instruction Manual page 9

Advertisement

Saniclave RS-SC-102
Warranty Registration
Please fill out this page, or a copy of it, and send it to the address below to validate
your warranty. Warranty Registration Card for the REVOLUTIONARY SCIENCE
Saniclave. (Please Print)
Date Purchased _____________________ Serial Number___________________
Purchased From___________________________________________________
(Please check one) ______ Company ________ School ________ Individual____
Name of Department or Institution:
_______________________________________________________________
_______________________________________________________________
Address:
_______________________________________________________________
_______________________________________________________________
_______________________________________________________________
_______________________________________________________________
Intended use:
_______________________________________________________________
Where did you hear of our product?
_______________________________________________________________
Comments:
_______________________________________________________________
_______________________________________________________________
_______________________________________________________________
_______________________________________________________________
_______________________________________________________________
Warranty
Your REVOLUTIONARY SCIENCE Saniclave is warranted to be free from defects of
material and workmanship under normal use for a period of two years from date of
purchase. If your Saniclave proves to be defective within this warranty period,
APR&D, Inc. will repair or replace your unit at our option. This warranty will be void
if malfunction is caused by accident, misuse, negligence, tampering, abuse, and/or dam-
age in transit. This warranty excludes any relief for incidental or consequential damag-
es.
R E V O L U T I O N A R Y S C I E N C E
10
Saniclave RS-SC-102
Operating Instructions
1. Place Saniclave on a dry, level surface.
2. Remove chamber door by rotating door counter clockwise.
3. Fill chamber with approximately 1 liter of distilled water, until the metal por-
tion of the probe is in contact with the water.
ALWAYS use distilled water. The sensor detects ion content in the water.
If distilled water is too pure, the autoclave may not detect the presence of
water. Salt may be added to the water to increase detectability. Class III
distilled water is recommended.
4. Place rack in the bottom of the chamber and place instruments to be sterilized
(unwrapped) on top of rack.
5. Inspect the door to assure that the seal is clean and not damaged. Be sure that
the inner portion of the door is also clean and clear, ensuring a secure fit.
6. If the silicone seal has been removed re-install silicone seal in the door by
evenly tucking seal under all lugs. Seal may be cleaned with warm, soapy water
(replace seal when it becomes noticeably cracked or worn).
7. Place the door on the chamber, ensuring that the arrow on the door is aligned
with the "open" arrow on the autoclave.
8. Press the door in snugly and gently turn the door handle clockwise until it can
no longer turn. The door arrow should be aligned with the "close" arrow on the
autoclave.
9. Insert the plug of the cord into the wall outlet. NEVER modify the plug to
make it fit, as the plug will only fit in one way. Autoclave must be grounded
when in use.
10. Standby Mode: The 3-digit LED indicator will turn on and appear as "-00."
11. Press 'START' to begin the 30 minute sterilization cycle.
12. Preheat Mode: Display will change to 'Ph' for the preheat mode (about 25
minutes).
13. Observe the display. If an error warning is displayed, the cycle will be aborted
(see troubleshooting section on page 10).
14. Sterilization Mode: After the preheat mode, the display will automatically
begin to count down from 30 minutes. For your convenience, the display will
show the temperature reading. To view the cycle time left, press 'START.' This
will not abort the cycle. It will change the reading on the display to cycle time.
15. To abort the cycle at any time unplug power cord. Display will then go blank
and chamber will instantly begin to slowly depressurize. Allow a few minutes for
the chamber to fully depressurize before restarting the autoclave.
16. Cycle Complete: "-CC" will display if the cycle was successful.
Caution: The chamber and contents may be hot to the touch.
R E V O L U T I O N A R Y S C I E N C E
7