Saniclave RS-SC-102
Biological Indicators
Microbial devices such as biological indicators (BI) are designed to accompany a load
to be sterilized. When a BI shows failure, the load must not be considered sterile.
Note: Wrapping, packing and loading improperly may result in failures. In addition, the
failure may be a result of sterilizer malfunction. Check load configuration and verify
that it is in accordance with the manufacturer's instructions. Also, check the BI's in-
structions for proper use, storage and disposal. Only BI's that are cleared by the FDA
and labeled for use with the same physical parameters (of exposure time and temper-
ature) of the Saniclave shall be used to monitor sterility of loads processed in Sani-
clave.
Class 5 Integrators
A Class 5 integrator are a type of chemical indicator. It does not contain microbial
organisms. A Class 5 Integrator monitors the time, temperature and pressure. These
single use devices are calibrated to confirm that parameters have been met. Only use
FDA cleared integrators that are validated for 121C for 30 minutes. Follow manufac-
turer's recommendations."
Use an FDA approved biological indicator (BI) weekly to confirm sterility and
STEAMPlus Class 5 integrators with each load. Follow the BI manufacturer's instruc-
tions. Revolutionary Science recommends the use of biological indicators and class 5
integrators manufactured by SPS Medical.
SPSmedical Supply Corp. 6789 W Henrietta Road Rush, NY 14543 USA (800) 722-
1529
www.spsmedical.com
R E V O L U T I O N A R Y S C I E N C E
16
Instruction Manual for the
Revolutionary Science Saniclave
Model RS-SC-102 and RS-SC-50
R E V O L U T I O N A R Y S C I E N C E
Makin g Scienc e Si mple

Table of Contents

Troubleshooting and Error Modes
Warranty, Registration and Policy
Contact Information
Checklist and Notes of Service Record
Care, Storage and Maintenance
Cycle Documentation Table
Sterilization Monitoring Guidelines
NOTE:
Read this manual before use!
safeguards and operating instructions for the operation of the REVOLUTIONARY
SCIENCE model Saniclave by Alternative Pioneering Research & Development, Inc.
(APR&D, Inc.)
© Alternative Pioneering Research and Development, Inc.
R E V O L U T I O N A R Y S C I E N C E
17319 Lake Blvd.
Shafer, MN 55074
Toll Free: (800) 775-7942
Fax: (775) 218-7015
2
3
4
4
5-6
6-7
8-9
10
11
11
12
13
14
15-16
This manual contains important information,
E-mail:
support@revsci.com
Website:
www.revsci.com
1

Hide quick links:

Advertisement

loading