KD2 Pro
Number of Measurements: This is the total number of
measurements that you want the KD2 Pro to make while in
Auto Mode. The number is between one and the amount of
available memory. As you change this value, the Experiment
Duration value changes. The KD2 pro stores up to 4,095 read-
ings.
Experiment Duration: This is the total amount of time that
your KD2 Pro operates in Auto Mode, according to your mea-
surement interval and number of measurements selected. This
gives you an idea of how long the total Auto Mode experiment
time takes.
Available Memory: This shows how much memory you have
available for storing measurements. The KD2 Pro Utility pre-
vents you from defining an experiment that could exceed the
amount of available memory. If you reach the limit of mem-
ory storage, the KD2 Pro stops taking measurements in Auto
Mode.
KD2 Pro Date/Time: This shows the current time in your
KD2 Pro. If the time is not correct, click the Cancel button,
then choose "Set KD2 Pro Date/Time..." from the Actions
Menu.
4. Click on the Start button. The KD2 Pro Utility then sends
your settings to the KD2 Pro.
5. Unplug the cable from the KD2 Pro, then connect the desired
thermal properties sensor to the instrument. Your KD2 Pro
is now ready to make autonomous measurements as you have
programmed it.
Viewing KD2 Pro Information
Choose "View KD2 Pro Information..." from the Actions Menu to
see information about your KD2 Pro.
4 THE KD2 PRO UTILITY
39
Need help?
Do you have a question about the KD2 Pro and is the answer not in the manual?