KD2 Pro
2 KD2 PRO OVERVIEW
The longer read times for the TR-1 and RK-1 sensors help to prevent
errors caused by effects from the large diameter needle and contact
resistance between the sensor and the sample in granular and solid
materials.
The shorter read time for the KS-1 sensor reduces the amount of
heat input into the sample. This is especially important with liquid
samples where excessive sample heating can cause errors from free
convection (Section 5.4) In low viscosity samples (e.g. water and
aqueous solutions), configure the KS-1 read time should be config-
ured to the minimum allowed one minute.
To set the read time, press menu and navigate to the Config menu.
Scroll down to Read Time and press Enter. Arrow down to the sen-
sor that you want and press Enter. Then press Enter again to scroll
through the read times for that sensor.