Error Indication - Philips PM 6303 Operating Manual

Rcl meter
Hide thumbs Also See for PM 6303:
Table of Contents

Advertisement

E
3-6
in
2
s
ion
20
so
loon
200
soo
ikn
2k
5 k
iokn
Zx
Fig.
6
Measurement
voltage
and
current
at
an
inductive
CUT (G>
10,
without
Rp,
with
Rp)
3.4.6.
Error indication
Several
functions
and
logical
states of
the
instrument
are
continuously
internally
checked
during
normal
operation.
Possible errors are indicated
by EO
...
E3 on
the
display.
The
meaning
of the
error
codes
are
given
in
the
following
table
Error
code
location of
malfunction
EO
RAM,
microprocessor
El
progr. amplifier,
level
detector
E2
counter
of
ADC,
integrator
control
section
E3
reference
measurement
circuitry
A
if
an
error
code
is
displayed
the
instrument should be switched
off.
If
after
switching
on
the
error
code
is
indicated again please
contact the
Philips service
organisation.
After switching
power
off
a
time
interval
of
at least
5
s
should
pass
by
-allowing the capacitors
of
the
power
supply
to
discharge-
before
the device
is
switched
on
again.
This procedure
is
necessary
to
set
the
internal logic circuitry
to
its
correct
initial
condition.

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents