Control Solution Results

The iPet PRO meter is functioning
correctly only if the control
solution test result is within the
specified range printed on the test
strip vial. If the test result is out of
range, please repeat the test.
Out of range result may be
caused by:
Incorrect steps of performing
the test
The control solution temperature
is lower than 59°F (15°C) or
higher than 95°F (35°C)
Expired or contaminated
control solution
Expired or contaminated
test strips
Meter malfunction
20
Note:
The result will not be included
in day average calculation when
the iPet PRO meter is set in the
control solution test
DO NOT use iPet PRO meter
if the problem persists. Please
contact customer service
immediately.
mode.