System Messages - RADIOMETER ABL 77 Series Service Manual

Blood gas and electrolyte
Table of Contents

Advertisement

Chapter 3: Troubleshooting
Introduction
This section lists system messages, provides an interpretation and any
recommended corrective action. This section groups system messages in
categories. Reference error messages using its unique error number.
System messages This table provides an interpretation of the message and suggests corrective action.
Error #
Message
1000
System Errors
1001
Battery is low
1002
Unable to allocate memory
1003
Calculation error
1004
The file can not be opened
1005
A file has been corrupted
1006
No data from analog board.
3-16

System Messages

Interpretation
Battery voltage is below
minimum charge
Insufficient memory
Unable to calculate
sensitivity values due to
impossible mathematical
values
Memory is unavailable.
The file being read has been
corrupted.
The analyzer does not
recognize any signal from
the analog board
ABL77 Service Manual
Corrective Action
Connect analyzer to A/C
power. If battery does not
re-charge refer to Chapter 4,
Test and Calibration
Procedures to test the
battery. Replace the battery
pack as necessary.
Cycle the analyzer power
off then back on to re-boot
the computer.
Repeat the measurement. If
this error persists,
investigate possible sources
of air contamination.
Shutdown the analyzer
power through the Main
Menu Shutdown button
then turn the analyzer back
on to re-boot the computer
Investigate a possible main
CPU failure
Check cable connections.
Replace the A/D ribbon
cable between the CPU and
analog board as necessary.
Secondarily it may be
necessary to replace the
Analog board.
Continued on next page