Chapter 4: Test and Calibration Procedures
Heater Circuit,
Verify the
Step
heater circuit
(continued)
4-38
Continued
5.
• If the Th reading is within the target range, the heater circuit is
functioning properly.
• If the Th reading is outside the target range, the sensor cassette may
be defective. Replace the sensor cassette and repeat step 4 and 5.
6.
If after replacing the sensor cassette the Th reading is outside the
target range, replace the analog board.