Vibby has an integrated Demo Mode built in, making multiple
simulations of a fall easier. Demo Mode is automatically entered after
exiting Storage Mode. Throughout this time the LED will continually
flash green. Demo Mode will continue running even after a fall has
been simulated.
1)
2)
When in Demo Mode, simulate a fall by fastening the wrist strap,
then hold the wrist strap with the Vibby facing down and at a
minimum of 60cm above the ground. Hold in this position for at least
30 seconds. Drop the Vibby avoiding it twisting so that it lands on its
face on the floor. Leave it on the floor for at least 20 seconds whilst it
analyses the fall. Once the fall has been detected the red LED will
begin to flash and the Vibby will begin to vibrate.. Leave the Vibby on
the floor and after a further period (20-30 seconds) the Vibby will
transmit the alarm; indicated by the red LED illuminating for a short
period.
The Vibby will automatically exit Demo Mode and enter Active Mode
after 5 minutes of exiting Storage Mode.