5 USING THE CONDUCTIVITY METER (continued)
4
Close the protection cover and press the
CAL button for over 2 seconds.
ash and the calibration value is
displayed. After the calibration is
completed,
illuminate steadily.
5
Clean the sensor with tap water and
remove moisture.
6
Press the MEAS button for 0.5 seconds to
enter the measurement mode and prepare
for measurement.
Note: The Elcometer 138 Conductivity Meter has a two-point calibration mode. For
details, see the full Elcometer 138 operating instructions which is available to
download via our website. The two-point calibration mode requires two standard
solutions.
5.6 TAKING A READING
1
Press the ON/OFF button to switch the
meter on.
2
Open the protection cover and place drops
of the sample on to the sensor.
Put an appropriate amount of the test sample
into the measurement cell avoiding the
inclusion of bubbles. Bubbles in the solution
may cause the conductivity measurement to be
inaccurate.
Cell
3
Close the protection cover.
Ambient air may cause the measurement values to fluctuate. To reduce
environmental interference, close the protection cover.
Without using the reading locking function:
4
Read the value displayed when
en-11
R
and
stop ashing and
CAL
Appropriate
Electrode
MEAS
and
CAL
Button
Too Little
appears.
www.elcometer.com
CAL
Button
CAL