5 USING THE CONDUCTIVITY METER (continued)
Conductivity S/cm Display
Press the CAL button
for 0.5 seconds
Conductivity S/m Display
5.5 ONE-POINT CALIBRATION
1
To set 'One-point calibration', switch the
meter on, press and hold the MEAS button
for 3 seconds. Press the CAL button until
appears.
CAL
2
Press the MEAS button again for 0.5
seconds and the calibration type will be
displayed (1 or 2). Press the CAL button
to change the setting to 1 as required and
press MEAS again to apply the setting.
3
Open the protection cover and place some
drops of the 1.41mS/cm standard solution
into the measurement cell. Washing the
sensor with standard solution beforehand
may provide more accurate calibration.
www.elcometer.com
Press the CAL button for
0.5 seconds
Press the CAL button for
0.5 seconds
TDS ppm Display
Press the CAL button
for 0.5 seconds
Salt Concentration % Display
MEAS
Button
R
CAL
Button
Protection
Cover
en-10