NOTICE
Risk of surface damage!
To prevent embrittlement of plastic parts, use only disinfectants with a low alcohol
•
content.
Do not use highly alkaline solutions or strong acids.
•
As a rule, be sure to use disinfectants in a way that no liquid can penetrate into the light
•
head or the suspension arm system.
The wipe disinfection procedure permitted by Gebrüder Martin can be performed with neutral
disinfectants (such as surface disinfectants) based on aldehydes, quaternary ammonium
compounds, etc.
The instructions of the disinfectant manufacturer regarding material compatibility, dosage and
exposure time must be duly observed.
Do not use substances causing stress cracks in acrylic glass or polycarbonate (such as
Surfanios or terralin
9.4
Machine Disinfection and Sterilization of the Handles
To enable sterile operation of the lights, the sensoGrip handle must be cleaned, disinfected and
sterilized prior to use.
NOTICE
Useful life of the handles!
The sterilizable handles are subject to wear by use and sterilization. Crack formation or
discoloration indicates material fatigue.
•
If any such signs are detected on the handle, do not resterilize it. Replace the handle
instead.
30
®
liquid).
Instructions for Use
Operating Lights marLED
®
V10/V16
V. 1.3