Fluke 415B Instruction Manual page 53

Table of Contents

Advertisement

415B
c.
Set the
POWER
and
HIGH
VOLTAGE
switches
to
ON.
Set the 500
volt
OUTPUT VOLTAGE
dial to
each setting
and
measure
the rectifier output.
(Differential
voltmeter
indications
are
via the
ten-
to-one
divider).
DIAL POSITION
VOLTS
DC
(±10%)
0
500
1000
1500
2500
1700
2300
2700
3500
4800
4-50.
DC
OUTPUT
RIPPLE.
This procedure
meas-
ures and displays the ac
ripple
superimposed
on
the
dc
output.
a.
Set
the
controls
of
the
Model
41
5B
as
follows:
POWER
POLARITY
OUTPUT VOLTAGE
dials
OFF
Negative
First
dial
to
2500;
second
dial to 500;
others
to
zero
b.
Connect a
0.
05
microfarad, 5000
volt,
capacitor
and
the
protection
circuit
shown
in
Figure 4-2
in
series between
the
OUTPUT
connector and
the input
of
a
preamplifier (Gain
=
100).
Connect
the
pre-
amplifier output
to the
input
of
an
oscilloscope.
c.
Set
the
POWER
and
HIGH
VOLTAGE
switches
to
ON
and observe
the
ripple.
The
ripple
should be
less than
1.0
millivolt
peak-to-peak
(see
Figure
4-5).
(Oscilloscope display should
indicate less
than
100 millivolts
peak-to-peak,
and are
typically
20
to
40 millivolts
peak-to-peak).
4-8
Rev.
1