Fluke 415B Instruction Manual page 47

Table of Contents

Advertisement

415B
OUTPUT VOLTAGE
Third
dial
to
20;
dials
fourth
dial
to
5;
other
dials
to
zero
HIGH
VOLTAGE
STDBY/RESET
b.
Connect
a
1,
000 ohm,
2
watt,
resistor
across
the
OUTPUT
connector.
c.
Set the
HIGH
VOLTAGE
switch
to
ON.
The
STDBY/
RESET
lamp
shall
extinguish
and
the
HIGH VOLT-
AGE
lanp
shall illuminate.
d.
Increase
the output voltage
in
1
volt step.
The
over-
load
circuit
should operate
and
extinguish
the
HIGH
VOLTAGE
lamp
within the
range
of
29
to 35 volts.
If
the trip
point
is
beyond
this
range, adjust
as
follows:
e.
Set the
HIGH
VOLTAGE
switch
to
STDBY/RESET.
f.
Set the
unit
for
maximum
current by turning
R5
fully
counter-clockwise.
This
control
is
located
on
the
High
Voltage
Rectifier
P/C
board and
is
ac-
cessible through a
vent
on
the left-hand side
of
the
unit.
g.
Set
the
OUTPUT VOLTAGE
dials
to
32
volts.
NOTE!
This voltage
will
provide
32
milliamperes
to
the
1,000
ohm
load
resistor.
If
it is
desired that
overcurrent
trip
should occur
at
a
lower
cur-
rent,
choose
the
output
voltage
to
provide
the
reduced
curren
t.
h.
Using an
insulated
screwdriver, slowly
turn
R5
clockwise
until the
overload
circuit trips
and
ex-
tinguishes
the
HIGH
VOLTAGE
lamp.
i.
Set
the
HIGH
VOLTAGE
switch
to
STDBY/RESET.
Set
the
OUTPUT VOLTAGE
dials
to
25
volts,
or
to
a
comparable
voltage
if
the
adjustment
was
for
an
overload current
less
than
32
milliamperes.
j.
Repeat
steps
c
and
d
to
check
the
adjustment.
4-34.
Meter
Calibration
a.
Set
the
controls
of
the
Model 415B
as
follows:
POWER
POLARITY
OUTPUT VOLTAGE
dials
HIGH
VOLTAGE
ON
Negative
First
dial to
2500;
second
dial to 500;
other
dials to
zero
STDBY/RESET
b.
Set
the
HIGH
VOLTAGE
switch
to
ON.
c.
Using an
insulated
screwdriver,
adjust
R26
on
the
Switch
P/C
board
so that the
panel
meter
indicates
3000
volts
(±90) volts.
NOTE!
It
will
be necessary
to
remove
the
cover
and
disable the
cover
interlock to
make
this
adjust-
ment.
d.
Set the
POLARITY
switch
to
Positive and check
the
+3000
volts indication.
4-35.
TROUBLESHOOTING
4-36.
Introduction
4-37.
The
information
in
the following
paragraphs
is
intended
to
assist
you
to isolate
most
common
troubles
in the
Model
41
5B
if
a
systematic approach
to
trouble-
shooting
is
followed.
A
good understanding
of
the
prin-
ciples
of
operation
is
necessary
to
efficient
trouble-
shooting,
and
it
is
recommended
that
you
review Section
IH before attempting
to
troubleshoot the instrument
in
detail.
4-38.
The
following troubleshooting
procedure
is
in
such
sequence
that
it
can be applied
to
any
unit,
in-
cluding
one
in
which
the trouble
is totally
unknown, and
there
is
doubt
whether power
can be
applied without
causing
damage.
If
the
unit is
operable, the
Resistance
Measurement, paragraph
4-40, and
the
Standby
Power
Check, paragraph
4-41,
may
be
omitted.
The
procedure
is in
two
parts:
Isolation
Checks, paragraph
4-39,
and
Assembly
Troubleshooting,
paragraph
4-45.
The
Iso-
lation
Checks
are designed
to
localize
most
common
troubles
as quickly as possible.
Certain key
circuit
parameters are checked
first,
which
may
eliminate a
large portion
of the
circuitry
as
a possible
cause
of
the
trouble.
If
the
desired
results
are
not
achieved
in the
isolation
check, reference
is
given
to the
appropriate
assembly
troubleshooting procedure.
NOTE!
When
troubleshooting the
unit,
use
negative
polarity
output.
Most
voltages
are
referenced
to
the positive
bus
which
is
grounded when
the
output
polarity
is
negative.
This
provides
some
protection
to
the
person doing
troubleshooting.
4-39.
isolation
Checks
4-40.
RESISTANCE
MEASUREMENTS.
These
checks
verify
correct resistance
between transformer
windings
and
between
the
shunting capacitors on the high voltage
transformer.
An
output
resistance
measurement
reveals
a possible short
circuit
in the output
and permits
a
check
of
sample
string
resistors before
the
unit is
turned
on.
a.
Connect an
ohm
meter
between
the yellow lead
on
the
first
OUTPUT VOLTAGE
dial
wiper
to
the jade
lead
on
the
transformer
terminal.
The
resistance
should increase
in
even
steps
from
approximately
360
ohms
to
1.
1
kilo
ohms
as
the
switch
is
stepped
up.
b.
Connect
the
ohmmeter
across
the front
panel
OUT-
PUT
connector.
The meter
should
indicate
approx-
imately
2
kilo
ohms
with
all
OUTPUT VOLTAGE
dials set to
zero and
the
POLARITY
switch
set
first
to
negative
and
then to
positive.
As
each
OUT-
PUT VOLTAGE
dial
is
stepped
up, the
resistance
should increase 500
ohms
per
volt
and
be approxi-
mately
1.
25
meg ohms
with
all
dials
to
maximum.
Rev, 2
4-5

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents