Fluke 415B Instruction Manual page 49

Table of Contents

Advertisement

41
5B
c.
Repeat
the
output
resistance
measurement
at the
rear
OUTPUT
connector.
With
ali
OUTPUT
VOLT-
AGE
dials
set to
zero,
the
resistance should
be
approximately
2
kilo
ohms.
4-41.
STANDBY POWER.
This procedure measures
power
consumption
in
the
STDBY/RESET
position.
It
reveals possible
gross
faults,
such as wiring errors
and shorted
components
in the
auxiliary
supply, 125
volt
regulator
and error
amplifier.
a.
Connect
the
Model 415B
through
a
Variac
to
a
115
volt,
60
Hz,
power
line
with a
wattmeter
or
an
ammeter
in
series with
the
ac
line.
Set
the
Variac
to
zero
output.
b.
Set
the
Model 415B
controls as follows:
OUTPUT VOLTAGE
dials
POWER
POLARITY
HIGH
VOLTA
GE
All
to
zero
ON
Negative
STDBY/RESET
d.
Slowly
increase
the
output
of
the
Variac.
The
STDBY/RESET
light
should
come
on approximately
60
seconds
after the
line
power
reaches
103
volts.
At
115
volts
ac
the
unit
should
draw
a
nominal
25
watts
(0.
36
amperes);
power consumption
should
be
no greater than
35 watts.
4-42.
125
VOLT SUPPLY
CHECK.
A measurement
of
ripple
on
the 125
volt
supply
at
low
and
high
line
voltage
reveals
quickly
not
only the ac
superimposed
on
the
dc
output but also the capability
of
the circuit
to
regulate.
The
value
of
the output voltage
should also be
measured.
NOTE!
The
following checks
must
be
done
with the
top cover
removed and
the interlock
disabled,
a.
Set the controls
of
the
Model 415B as
follows:
OUTPUT VOLTAGE
All
to
zero
dials
HIGH
VOLTAGE
STDBY/RESET
POWER
ON
b.
Connect
an oscilloscope
between
the top
of
R8
on
the
Amplifier
P/C Assembly
and chassis
ground.
Adjust
the
Variac
output
between
103
and
127
volts.
The
ripple
should be less
0.
8
millivolts
peak-to-
peak
(see
Figure
4-4).
c.
Connect
a differential
voltmeter between
R8
and
chassis ground.
At
115
volts
ac
line,
the
dc
volt-
age should be
between
120
and 130
volts.
Over
a
line
voltage
range
of
103
to
127
volts, the
dc
output
should
vary
less than
2. 5
millivolts
from
the in-
dication
at
115
volts.
d
-
the
125
volt
supply
fails
to
regulate or
the value
of
output voltage
is
incorrect, refer
to
paragraph
4-46.
5
MS/CM
Figure
4-4.
R IPPLE
ON
1
25
VOLT
SUPPLY
4-43.
ERROR
AMPLIFIER CHECK.
This
procedure
determines quickly whether
the
Error
Amplifier
is
operational
and provides
sufficient gain
to
drive
the
grids
of
the
Series
Pass
tubes.
a.
Set the
Variac
output
to
115
volts.
Set the
Model
415B
controls
as
follows:
OUTPUT VOLTAGE
dials
HIGH
VOLTAGE
POLARITY
POWER
One-volt
dial to
one
volt; all
others
to
zero
STDBY/RESET
Negative
ON
b.
Connect a
VTVM
between
either
end
of
R8
(or
Rll)
on
the
Series Pass
P/C
Assembly
and chassis
ground.
The
voltmeter should
indicate
approxi-
mately one
volt positive.
c.
Connect a
battery
of
1.
5
volts
or greater
to
the
OUT-
PUT
connector:
positive to positive
and negative
to
negative.
When
the
battery voltage
is
applied,
the.
VTVM
at the
Series
Pass
tube grid should
swing
to
approximately
100
volts
negative (between -90
and
-120
volts).
d.
The
100-volt
change
in
step
c
indicates that
the
auxiliary
power
supply voltages and
the
Error
Amplifier are probably
all
right.
If
amplifier
operation
is
unsatisfactory, refer to
paragraph 4-47.
4-44.
TURN
ON.
This procedure checks
out the
High
Voltage Rectifier
and
the
Series
Pass
tubes.
a.
Disconnect
the battery
across
the
OUTPUT
con-
nector.
Set
the
first
OUTPUT VOLTAGE
dial
to 500.
b.
Set the
HIGH
VOLTAGE
switch
to
ON.
The
line
ac
wattmeter/ammeter
should
indicate
approxi-
mately
38 watts
(0.
52
amperes).
The
front
panel
meter
should
indicate
approximately
500
volts.
CAUTION!
If
the panel
meter
indicates greater
than
600
volts,
immediately
set
the
HIGH
VOLTAGE
switch
to
STDBY/RESET.
If
the
isolation
pro-
cedure has
been
followed thus
far, a
possible
cause of
trouble
is
a
shorted diode
in
the
High
Voltage
Rectifier.
4-6
Rev.
1

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents