CCESSORIES
Performance: Operating pressure range: 4 to 20 cm H
Dimensions (H x W x D): 4.4" x 6.5" x 5.7" (112 mm x 164 mm x 145 mm)
Weight: 2.9 lb (1.3 kg)
Power Supply
Input range for
2.5 A <140 VA (110 W) (maximum power consumption). Instantaneous peak
power consumption <340 VA.
Refer to the DC-12 converter instructions for DC ratings. Only use the ResMed
DC-12 converter for DC input.
Actual power consumption will vary, depending on factors such as the use of
accessories, height above sea level, and ambient temperature. Power consumption
values for typical treatment conditions (using a breathing machine set to 0.5 L and
15 breaths/min with 6'6" (2 m) air tubing, and an
following table
Device
S8 Elite
Power outlet for HumidAire 3i
Outlet range: 100–240V, 50/60Hz; 110V, 400Hz; 2.0A
Air Filter: Two-layered, powder-bonded, polyester non-woven fiber
Air Tubing: Flexible plastic, 6'6" (2 m)
Air Outlet: The 22 mm conical air outlet complies with EN 1281-1
IEC 60601-1 Classifications: Class II (double insulation), Type CF
Noise level: Radiated sound pressure is measured at <30 dBA at 10 cm H
1 m (3'3.37") according to ISO17510
A
CCESSORIES
D
C
ATA
ARD
Dimensions (L x W x D): 2.56" x 0.87" x 0.03" (65.04 mm x 22.15 mm x 0.85 mm)
)
with
S8 E
H
A
LITE
UMID
IRE
Pressure
(cm H
O)
2
5
10
15
20
O
2
: 100–240 V, 50/60 Hz; 110 V, 400 Hz;
3i
mask) are given in the
U
M
LTRA
IRAGE
Power
Power
(VA)
(W)
18
7.6
23
11.6
30
15.9
39
20.7
S
YSTEM
O and
2
37
S
PECIFICATIONS