Using Status Reporting Registers - Omega Engineering OMB-NETSCAN 1500 User Manual

Ethernet/internet-based data logging & control instrument
Table of Contents

Advertisement

Registers, Data Formats, & Queries
Bit
Description
Location
DI01
Alarm
Trigger
DI02
Detected
DI03
Ready
DI04
Scan
Available
Message
DI05
Available
(MAV)
DI06
Event Status
Register Bit
(ESB)
Buffer
DI08
Overrun

Using Status Reporting Registers

As mentioned earlier the status reporting registers are organized in a hierarchical structure with the lower level
registers containing more general information and the higher level registers containing more detailed
information about particular events. In general, the lowest level register, the Status Byte Register (STB)
contains information that may require more immediate action from the controller in that events contained in this
register tend to be more time critical.
However, as mentioned earlier, other registers may access the STB via the ESB bit. This allows any condition
within the status reporting register hierarchy to have access to the STB and therefore Serial Poll and Service
Request functionality. The following is an example to show, in brief, how the status reporting register hierarchy
works. Notice that the clearing of the higher level registers cleared the bits in the lower level registers that were
associated with the root cause of the condition.
PRINT#1,"N0 X N16 X"
.
PRINT#1,"E?X"
INPUT#1,A$
PRINT A$
PRINT#1,"U2X"
INPUT#1, A$
PRINT A$
PRINT#1,"U0X"
INPUT#1, A$
PRINT A$
PRINT#1,"U1X"
INPUT#1, A$
PRINT A$
D-16
Set, Clear and Descriptive Information
Set when the scanning device has sensed a valid alarm condition. The alarm is
cleared when the alarm condition no longer exists.
Set when the NetScan has detected the trigger condition and will be cleared when the
acquisition is complete or the trigger has been reconfigured.
Set when the NetScan is ready to process another command. It is cleared when the
NetScan is processing a command line. This bit should be examined with a serial poll
prior to issuing a new command line. This allows any detected errors to be traced to
the specific command line containing the error. If all the setup information for a
specific NetScan operation is included in one line, this bit also indicates when all
processing is done and the X command is completed. This ensures that the NetScan
is done processing all state changes before initiating any further activity.
Set when at least 1 acquisition scan is available in the acquisition buffer to be read.
Cleared when there are no scans available in the buffer to be read.
Set when the output queue is not empty. It is cleared when the output queue is empty.
This bit reflects whether any command responses are still in the output queue.
Reflects the logical OR of all the bits in the Event Status Register (ESR) ANDed with
their equivalent enable bits in the Event Status Enable (ESE) register. If this bit is set,
at least one bit in the ESR is set and has its corresponding enable bit in the ESE set.
The status command U0 can be issued to read the ESR. See the following for more
information on ESR and ESE.
Set if a buffer overrun occurs. It is cleared when the buffer becomes empty by either
reading out the contents of the buffer or performing a Reset (*B) of the buffer.
' Configure ESB in the STB to be set when a Device Dependent
' Error occurs.
' Cause an error.
' Query the ESC to determine the cause of the error
' Get the response
' Screen shows E016.This is a calibration error
' Query the CSR to determine the cause of the ' calibration error
' Get the response
' Screen shows E002. This is a calibration Gain error
' Now query the ESR (reading ESC should have cleared it)
' Get the response
' Screen shows 000 — Device Dependent Error no longer there
' Now query the STB.
' Get the response
' Screen shows 4 — ESB bit no longer set (only Ready is set).
Appendix D
NetScan User's Manual

Advertisement

Table of Contents
loading

This manual is also suitable for:

Omb-netscan