CSN14/LV/ (T, B, & S)
There are three versions of low voltage signal conditioning cards for use with NetScan. Each card supports 16
differential input channels and is capable of measuring analog input signals on any of four programmable ranges:
±100 mV, ±1 volt, ±5 volt and ±10 volt. The cards share the same signal conditioning characteristics, but differ in
their style of input connector, i.e., terminal strip, BNC, and safety jack (CSN14/LV/T, CSN14/LV/B and
CSN14/LV/S, respectively).
Number of Inputs: 16 differential; programmable by channel for input range
Input Types: ±10V, ±5V, ±1V, and ±100mV
Input Connectors: Removable screw terminal blocks, BNC, or Safety Jacks
Voltage Range, Accuracy
± 100 mV;
± 1 V;
± 5 V;
± 10 V;
Digital Filtering:
Used for AC line cycle noise rejection.
Selectable off, 1-256 samples in a binary count.
Temperature Coefficient:
< (0.1 x rated accuracy)%/°C
Input Impedance:
Input Bias Current: 20 nA maximum
Common Mode Rejection: 100 dB typical
Maximum Common Mode Voltage:
Maximum Allowable Input: ± 25V rms
Channel-to-Power Ground Isolation: 200 V peak
Channel-to-Channel Isolation: 200V peak
7-14
Low Voltage Cards
1
and Resolution:
± 0.02%;
3.05 µV/bit
± 0.02%;
30.5 µV/bit
± 0.02%;
153 µV/bit
± 0.02%;
306 µV/bit
1MΩ typical
2
200 VDC, or 200 VAC peak
NetScan User's Manual